Main Specifications CA-2000

Model

 

 

CA-2000S

CA-2000W

 

 

CA-2000T

 

Light receptor

 

CCD image sensor (monochrome); 2/3-inch; Effective number of pixels: 1,000 x 1,000 pixels; Equipped with XYZ filter (closely matches CIE 1931 color-

 

matching function) and ND filter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lens

 

 

Interchangeable

Standard, wide, and telephoto lenses; low-magnification and high-magnification macro rings (for use with telephoto lens)

Measurement points (Resolution)

980 x 980 (Available to select 490 x 490 or 196 x 196 by using Data Management Software CA-S20w)

 

Color indication modes

XYZ, Lvxy, Lvu’v’, T uv, Dominant wavelength, Excitation purity, Lv contrast

 

 

 

Display modes

 

Pseudocolor, RGB image, Chromaticity diagram, Spot, 3D graph, Histgram, Nonuniformity enhancement, Cross section, Color difference, Multi-screen

 

 

 

 

Standard lens

Wide lens

Telephoto lens

 

 

Measurement sizes

With low-magnification macro ring

With high-magnification macro ring

 

 

 

 

 

 

 

(length per side of square) (*1)

Approx. 98 mm or more

Approx. 145 mm or more

Approx. 115 mm or more

Approx. 57mm (Fixed)

Approx. 27mm (Fixed)

 

 

 

 

(depending on the distance)

(depending on the distance)

(depending on the distance)

 

 

 

 

 

 

 

 

 

 

98 mm / 250 mm mm Approx.

145 mm / 200 mm

Approx.

115 mm / 900 mm Approx.

 

 

Measurable size for typical measure-

210 mm / 500 mm Approx.

410 mm / 500 mm

Approx.

275 mm / 2,000 mm

Approx.

57 mm / 500 mm Approx.

27 mm / 300 mm Approx.

ment distances (size/distance)

440 mm / 1,000 mm

Approx.

850 mm / 1,000 mm

Approx.

420 mm / 3,000 mm

Approx.

(Fixed)

(Fixed)

 

 

 

 

890 mm / 2,000 mm

Approx.

1,770 mm/ 2,000 mm

Approx.

 

 

 

 

 

 

 

 

 

 

Measurement luminance range

0.1 - 100,000 cd/m2

0.1 - 100,000 cd/m2

0.5 - 100,000 cd/m2

0.5 - 100,000 cd/m2

1 - 100,000 cd/m2

(including ND filter use)

Measurement time (*2)

Single : Approx. 5 sec. or more;

4-time integration: Approx. 6

sec. or more; 16-time integration: Approx. 10 sec. or more; 64-time integration : Approx. 28 sec.

or more; 256-time integration : Approx. 98 sec. or more

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Luminance

±3 %

 

±3 %

 

 

±3 %

 

±3 %

±3 %

 

 

 

Chromaticity

±0.005

 

±0.005

 

 

±0.005

 

±0.005

±0.005

Accuracy (*3)

 

Temperature/humidity drift (within the operating temperature/humidity range)

 

 

 

 

 

 

 

Luminance: ±2% of change compared to reference temperature and relative humidity of 23°C and 40%

 

 

 

 

 

Chromaticity: ±0.004 of change compared to reference temperature and relative humidity of 23°C and 40%

 

Repeatability (*4)

Luminance

0.5 %

 

0.5 %

 

 

0.5 %

 

0.5 %

0.5 %

 

 

 

Chromaticity

0.001

 

0.001

 

 

0.001

 

0.001

0.001

 

 

 

Luminance (*6)

±2 %

 

±2 %

 

 

±2 %

 

±2 %

±2 %

Inter-point error Chromaticity (*6)

±0.002

 

±0.002

 

 

±0.002

 

±0.002

±0.002

(*5)

 

 

Luminance (*7)

±3 %

 

±3 %

 

 

±3 %

 

±3 %

±3 %

 

 

 

Chromaticity (*7)

±0.003

 

±0.003

 

 

±0.003

 

±0.003

±0.003

Other functions

 

Interval measurement, Measurement sync (Synchronization frequency selectable), Integration function, Enhanced nonuniformity display, Conversion of en-

 

hanced nonuniformity image into numerical values (statistical values, etc.), Pixel binning function

 

 

 

 

 

 

Interface

 

 

USB 2.0 or higher

 

 

 

 

 

 

 

 

Operating temperature and humidity range (*8)

10-30°C, Relative humidity 70% or less/No condensation

 

 

 

 

Storage temperature and humidity range (*8)

0-30°C, Relative humidity 70% or less/No condensation, 30-35°C, Relative humidity 55% or less/No condensation

 

 

 

 

 

 

 

 

Size

 

Body only

160 (W) × 164 (H) × 199 (D) mm (Height including handle: 211 mm)

 

 

 

 

When lens and lens hood are attached

224 (D) mm

 

219 (D) mm

 

 

224 (D) mm

 

230 (D) mm

237 (D) mm

 

 

 

 

 

 

Weight

 

 

3.5 kg approx. (when standard

lens and lens hood are attached)

 

 

 

 

Power source

 

AC adapter 100-240 V , 1.2 A, 50-60 Hz

 

 

 

 

 

 

 

 

 

Lens Hood

CA-A68

 

CA-A69

 

 

CA-A68

 

CA-A70

 

Standard

 

Hood Cap

CA-A71

 

CA-A75

 

 

CA-A71

 

 

 

accessories

Calibration data DVD

Included with each lens.

 

 

 

 

 

 

 

 

 

 

Other

Mount Cover CA-A61, AC Adapter, USB Cable IF-A18, Data Management Software CA-S20w

 

 

Optional accessories

Soft Case CA-A60, Tripod CS-A3, Pan Head CS-A4, Calibration certificate

 

 

 

*1: Error in angle of view: 7%

*2: Measurement time differs depending on brightness of measurement object, PC performance, and data processing contents.

The specifications above were obtained under Konica Minolta's measurement conditions shown below:

PC's CPU

:

Pentium 4, 2.8GHz

Data processing

: Pseudocolor display of luminance data

Resolution

:

490 x 490

Shutter speed

:

Y measurement : 1/64 sec., XZ measurement : 1/32 sec.

Measurement subject brightness: Standard/wide lens: Approx. 80 cd/m2, Telephoto lens: Approx. 300 cd/m2 Low-magnification macro ring and telephoto lens: Approx. 400 cd/m2, High-magnification macro ring and telephoto lens: Approx. 600 cd/m2

*The measurement time becomes longer when the object is dark. The longest measurement time is approx. 10 seconds with 1-time integration, approx. 27 seconds with 4-time integration, approx. 95 seconds with 16-time integration, approx. 6 minutes and 8 seconds with 64-time integration and approx. 24 minutes and 19 seconds with 256-time integration

*3: The specifications above were obtained under Konica Minolta's measurement conditions shown below: Measurement subject brightness: Standard/wide lens: Approx. 35 cd/m2, Telephoto lens: Approx. 140 cd/ m2

Low-magnification macro ring and telephoto lens: Approx. 250 cd/m2,

High-magnification macro ring and telephoto lens: Approx. 250 cd/m2

Distance: Minimum distance of each lens, Evaluation: Based on the average obtained within 10% range at the

center of the screen, Temperature: 23°C±2°C, Relative humidity: 40%±10%, Measuring light: White, reference

light source, Integration: 64 times (Normal mode)

*4: The specifications above were obtained under Konica Minolta's measurement conditions shown below:

Resolution: 196 x 196, Shutter speed: Y measurement: 1/64 sec., XZ measurement: 1/32 sec. G a i n : N o r m a l

(x1), Light intensity level: Midpoint of appropriate light intensity range, Evaluation: Based on the maximum repeat-

ability (2 ) of all pixels, Temperature: 23°C±2°C, Relative humidity: 40%±10% , Measurement subject:

White, reference light source, Integration: 64 times (Normal mode)

*5: The specifications above were obtained under Konica Minolta's measurement conditions shown below: Measurement subject brightness: Standard/wide lens: Approx. 40 cd/m2, Telephoto lens: Approx. 160 cd/m2

Low-magnification macro ring and telephoto lens: Approx. 200 cd/m2,

High-magnification macro ring and telephoto lens: Approx. 350 cd/m2

Distance: Calibration distance of each lens, Resolution: 196 x 196

Evaluation (*6) : Based on the maximum/minimum values obtained in a square at the center of the screen measuring 60% of the height and width of the entire screen

(*7) : Based on the maximum/minimum values obtained in the entire screen

Temperature: 23°C±2°C, Relative humidity: 40%±10%, Measurement subject: White, reference light source,

Integration: 64 times (Normal mode)

*8: Even if the instrument is stored within the specified usage (or storage) temperature humidity range, the displayed value may change depending on long-period usage or storage conditions.

SAFETY PRECAUTIONS

For correct use and for your safety, be sure to read the instruction manual before using the instrument.

Always connect the instrument to the specified power supply voltage. Improper connection may cause a fire or electric shock.

CA-S20w System Requirements

OS

Windows XP Professional 32-bit SP3, 64-bit SP2; Windows Vista Business 32-bit, 64-bit

 

Windows 7 Professional 32-bit, 64-bit

 

(English, Japanese and Korean versions

 

For Windows® XP professional 64-bit, English and Japanese versions only)

CPU

Pentium® 4 2.8 GHz equivalent or higher

Memory

1024 MB or more

Hard Disk

Needs free space of 80 MB at least on system drive (where OS is installed)

 

In addition, each lens needs the following free spaces for installing calibration data.

 

For standard lens: approx. 540 MB

 

For wide lens: approx. 470 MB

 

For telephoto lens: approx. 1.3 GB

 

Also to save measurement data on hard disk, additional free space is required.

 

(approx. 11 GB minimum for 1000 measurement data in resolution of 980 x 980)

Display

Display capable of at least 1280 x 1024 dots / High color, 16-bit (Full color, 32-bit recommended)

Others

Optical drive capable of reading CD-ROM (for installing software) and DVD-ROM (for installing calibration

 

data) necessary.

 

USB port: USB ver. 2.0; Type A connector; For connecting measuring instrument

 

Excel® 2003 (under Windows® XP), Excel® 2007 (under Windows® Vista/Windows® 7), or

 

Excel® 2010 (under Windows® 7) necessary for continuous measurements using automation.

ssWindows®, Excel® is a registered trademark or a trademark of Microsoft Corporation in the United States and other countries. ss0ENTIUM® is a registered trademark or a trademark of Intel Corporation in the United States and other countries.

The specifications and drawings given here are subject to change without prior notice.

-If you have any questions about specifications, please contact your Konica Minolta representative.

-The specifications given here are subject to change without prior notice.