TIC-LF494B

Separate type LF490/LF612 and LF494/LF612F flowmeters

Terminal board

Thick walled steel conduit

Signal cable

(2-wire shielded hard-rubber sheathed cable)

Connected detector

Excitation cable (3-wire shielded hard-rubber cable)

IV wire

5.5 mm2 or more

Grounded with 100Ω or less ground resistance

Power switch

(External double-pole power switch )

[Instrument panel : ordered separately]

Grounded with 100Ω or less ground resistance

Power supply

Current output (4~20mAdc) or PROFIBUS

Digital output 1

Signal common for DI and DO

Digital output 2

Digital input (20~30Vdc)

Power cable

Digital input cable

 

Grounded with 100Ω or

(CVV)

(CVV-S)

 

 

 

 

less ground resistance

 

 

 

Figure 10. Separate type LF490/LF612 and LF494/LF612F flowmeters wiring Diagram

Table 1. LF610, LF610F, LF612 and LF612F Converters Signal Table

Symbol

Description

Cable

L1 ()

Power supply

Power cable (CVV)

L2 ()

 

 

GND

Ground (for arrester)

 

FG

Frame ground

 

DI

Digital Input (2030Vdc)

 

DO1

Digital Output 1

 

DO2

Digital Output 2

I/O cable (CVV-S)

COM

Signal Common for DI, DO1, DO2

 

 

 

Current Output (420mAdc)

 

 

Shielded cable for

 

or PROFIBUS

 

PROFIBUS-PA

 

 

 

X

 

Excitation cable

Y

Excitation Output

(for LF612, LF612F only)

E

 

 

 

A

 

Signal cable

B

Signal Input

(for LF612, LF612F only)

G

 

 

 

9

Page 9
Image 9
Toshiba TIC-LF494B manual Separate type LF490/LF612 and LF494/LF612F flowmeters