UNIVERSAL ANALYZERS MODEL 510S SAMPLE COOLER
SPECIFICATIONS |
|
SAMPLE FLOW RATE: | 0 to 2 L/M (at STP) |
MAXIMUM INLET TEMPERATURE: | 500o F. (260o C.) |
MAXIMUM INLET GAS DEW POINT: | 180o F. (82o C.) |
MAXIMUM INLET WATER CONCENTRATION: | 50%* |
MINIMUM AMBIENT TEMPERATURE: | 32o F. (0o C.) |
MAXIMUM AMBIENT TEMPERATURE: | 105o F. (41o C.)* |
MAXIMUM COOLING POWER: | 50 BTU'S/Hr. (47 kJ/Hr.) |
OUTLET SAMPLE DEW POINT: | 41o F. (5o C.), adjustable |
MAXIMUM INPUT POWER: | 150 WATTS |
VOLTAGE: | 90 - 125 VAC, 50/60 Hz |
or | 180 - 250 VAC, 50/60 Hz |
| |
ELECTRICAL CLASSIFICATION: | NEMA 1 |
DIMENSIONS: | 9" x 10" x 6", HWD |
WEIGHT: | 11 LB's., (5 KG) |
*at reduced flow rates. See capacity chart.
See us on the WEB at http://www.universalanalyzers.com