Hearing HelperPersonal FM

Hearing HelperPersonal FM Receiver • Model PFM R31

Acoustic Specifications:

ANSI 2.44 Test

Receiver • Model

FM Response (PFM R31)

Input:

80 dB random noise

 

 

Device:

KEMAR mannequin with BTK 3550 tester

Y:

80.0 dB 80 db

Main Y:

34.1 dB

X:

100 Hz + 8 OCT LOG

Main X:

4 kHz

ANSI S3.22–1987 Test

FM Response (PFM R31)

Minimum Low Cut (Tone Set to Low)

Environmental Mic Response (PFM R31)

Input:

80 dB random noise

 

 

Device:

KEMAR mannequin with BTK 3550 tester

Y:

80.0 dB 80 db

Main Y:

50.2 dB

X:

100 Hz + 8 OCT LOG

Main X:

4 kHz

Maximum Low Cut (Tone Set to Hi)

PFM R31

Max. SSPL90:

135.8 dB at 500 Hz

Max. SSPL90:

133.5 dB at 1000 Hz

HF Avg. SSPL90:

127.1 dB

HF Avg. SSPL90:

127.0 dB

HF Avg. Full On Gain:

37.7 dB at 60 dB in

HF Avg. Full On Gain:

35.4 dB at 60 dB in

Reference Test Gain:

38.1 dB

Reference Test Gain:

35.8 dB

Response Limit:

80.1 dB

Response Limit:

78.01 dB

 

F1=200 Hz, F2=8 kHz

 

F1=200 Hz, F2=8 kHz

Total Harmonic Dist:

3.6 % at 500 Hz

Total Harmonic Dist:

1.1 % at 500 Hz

 

3.9 % at 800 Hz

 

2.7 % at 800 Hz

 

2.2 % at 1600 Hz

 

2.1 % at 1600 Hz

 

 

 

 

Medium Low Cut (Tone Set to Mid)

 

Environmental Mic Response (PFM R31)

Max. SSPL90:

135.5 dB at 600 Hz

No Tone Modification

 

HF Avg. SSPL90:

127.1 dB

Max. SSPL90:

134.8 dB at 500 Hz

HF Avg. Full On Gain:

36.8 dB at 60 dB in

HF Avg. SSPL90:

126.4 dB

Reference Test Gain:

37.3 dB

HF Avg. Full On Gain:

40.9 dB at 60 dB in

Response Limit:

79.4 dB

Reference Test Gain:

40.8 dB

 

F1=200 Hz, F2=8 kHz

Response Limit:

81.1 dB

Total Harmonic Dist:

3.4 % at 500 Hz

 

F1=200 Hz, F2=8 kHz

 

3.4 % at 800 Hz

Total Harmonic Dist:

1.1 % at 500 Hz

 

2.2 % at 1600 Hz

 

1.1 % at 800 Hz

 

 

 

0.3 % at 1600 Hz

NOTE: SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE!

 

©2007, Williams Sound Corp. MCAT 005B

2

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 2
Image 2
Williams Sound PFM R31 specifications Hearing Helper Personal FM, Receiver Model, Acoustic Specifications Ansi 2.44 Test