Operation

Power Supply Operation

Configure The DISPLAY CONFIG menu option allows you to select the readback values Display displayed when the power supply is operating in its default state.

The factory default is to display voltage and current readback, but you can also choose voltage and power, current and power, or voltage, current, and power.

This table shows how to access and work with the DISPLAY CONFIG option.

Step #

Do This

You Will See

1

MENU

ERROR MSGS

 

 

 

 

 

2

until you see the “DISPLAY

DISPLAY CFG

 

CFG” option1

 

3

ENTER

Show V, I & P

 

 

 

 

 

4

Using the arrow keys, select

Show V and I

 

from Show V and I, Show V and P,

 

 

Show I and P, Show V, I, & P2.

 

5

ENTER

The setting is saved.

 

1.You can also press MENU repeatedly to bring up the Display Config option.

2.The option, Show V, I, & P is not available in current share mode.

Lock Out The KNOB LOCKOUT menu option allows you to lock the front panel knobs, Control forcing changes to be made via the VOLTAGE and CURRENT keys. Knobs should Knobs be locked out whenever you do not want someone to accidentally adjust the supply

settings while the unit is operating in local mode.

To lock out both knobs:

Step # Do This

 

You Will See

1

MENU

 

ERROR MSGS

 

 

 

 

 

 

 

2

 

until you see the “KNOB

KNOB LOCKOUT

 

 

LOCKOUT” option

 

3

ENTER

 

Lock V Knob? N

 

 

 

 

 

 

 

4

 

 

Lock V Knob? Y

5 ENTER

6

7ENTER

Lock I Knob? N

The Voltage knob is locked out.

Lock I Knob? Y

The setting is saved and the display returns to its default operating mode.

98

Operating Manual for XDC Series Power Supply

Page 100
Image 100
Xantrex Technology XDC manual To lock out both knobs