Calibration

Exit calibration mode

Exit calibration mode

When you have completed calibration, hit the EXIT key.

The SCPI command is:

CAL:STAT OFF, "0000"

Restore Factory Calibration

To restore the unit to the calibration constants set at the factory:

Step #

Do This

You Will See

1

 

Output V Cal

2

or

Factory Cal

 

Scroll to select Factory Cal.

 

3

Select Y to Restore the factory

Restore? Y

 

calibration. Selecting N will exit.

 

The SCPI command is

CALibration:RESTore

Note This procedure should not be used in place of regular calibration, but may be useful to restore the unit to an operational state in case of failure.