Omega DP26-TC manual Specifications

Page 57

SECTION 13. SPECIFICATIONS (continued)

ALARM OUTPUTS

2 Form "C" on/off relays. Configurable for latched and

(if applicable)

unlatched by software.

 

 

Max current: 5 AMPS, resistive load

 

 

Max voltage: 250 Vac or 30 Vdc

ENVIRONMENT

 

~ AC units

115/230 V~(AC) + 10%, 50/60 Hz

 

 

7 W max, power consumption (Non-Isolated Analog Out)

 

 

8 W max, power consumption (Isolated Analog Out)

 

DC units

10-32 Vdc or 26-56 Vdc,

 

 

 

6 W max, power consumption (Non-Isolated Analog Out)

 

 

7 W max, power consumption (Isolated Analog Out)

External Fuse Required:

 

 

 

 

 

IEC 127-2/III

 

 

 

 

 

 

Power

 

Fuse

 

115 V

 

125 mA @ 250 (T)

 

 

 

 

 

230 V

 

63 mA @ 250 (T)

 

UL 248-14 (Listed Fuse)

 

 

 

 

Power

 

Fuse

 

115 V

 

175 mA @ 250 V Slow-Blow

 

230 V

 

80 mA @ 250 V Slow-Blow

 

 

 

 

ENVIRONMENT

 

Operating temperature:

0° to 50°C (32° to 122°F)

 

(115/230 V rms + 10%)

Storage temperature:

-40° to 85°C (-40° to 185° F)

Relative humidity:

90% at 40°C (non-condensing)

MECHANICAL

 

Panel cutout:

1/8 DIN 3.62 x 1.78" (45 x 92mm)

Weight:

1.27 lb (575 g)

Case material:

Polycarbonate, 94 V-O UL rated

Protection:

NEMA-4/Type 4 Front BezeL

49

Image 57
Contents User’s Guide Servicing North America Page Page Page Page Page Page Introduction Safety Considerations About the Meter Page Page Page Page Page Back of the Meter Page Page Page Page Page Panel Mounting Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Page Specifications Side View Factory Preset Values CE Approvals Information WARRANTY/DISCLAIMER Shop on line at omega.com