Orion XX14I Aligning the Secondary Mirror, Adjusting the Secondary Mirror’s Axial Position

Page 21

Figure 37. This image shows the XX14i properly set up for collimation. Note the white paper placed across from the focuser and the horizontal orientation of the optical tube. Ideally, the telescope should be pointing at a light-colored wall.

and causing damage if something comes loose while you are making adjustments. Place a sheet of white paper inside the optical tube directly opposite the focuser. The paper will provide a bright “background” when viewing into the focuser. When properly set up for collimation, your telescope should resemble Figure 37.

Aligning the Secondary Mirror

To adjust the secondary mirror collimation, you will need a Phillips screwdriver and a 2mm hex key, or Allen wrench.

You will need to check, and adjust if necessary, four aspects of the secondary mirror’s alignment:

1.The secondary mirror’s axial position

2.The secondary mirror’s radial position

3.The secondary mirror’s rotational position

4.The secondary mirror’s tilt

The first three will probably only need to be checked and (pos- sibly) adjusted once. Thereafter, it is usually only the second- ary mirror’s tilt that will need to be adjusted occasionally.

Adjusting the Secondary Mirror’s Axial Position

With the collimating cap in place, look through the hole in the cap at the secondary (diagonal) mirror. Ignore the reflections for the time being. The secondary mirror itself should be cen- tered in the focuser drawtube. If it is off-center along the axis of the telescope, i.e., positioned too far toward the front open- ing or toward the rear of the telescope, as it is in Figure 35b, you will have to adjust the mirror’s axial position.

To do so, use the 2mm hex key to loosen the three small alignment set screws in the center hub of the 4-vane spi- der several turns. Now, grasp the mirror holder (the cylinder that is attached to the back of the secondary mirror itself) with one hand while turning the center screw with a Phillips head screwdriver with your other hand (Figure 38). Turning the screw clockwise will move the secondary mirror toward the front opening of the optical tube, while turning the screw counter-clockwise will move the secondary mirror toward the

Figure 38. To center the secondary mirror axially under the focuser, hold the mirror holder in place with one hand while adjusting the center screw with a Phillips screwdriver.

Spider vane thumb nuts

Figure 39. To center the secondary mirror radially in the focuser drawtube, make adjustments to the two knurled spider vane thumb nuts that are perpendicular to the focuser.

primary mirror. When the secondary mirror is centered axially in the focuser drawtube, rotate the secondary mirror holder until the reflection of the primary mirror is as centered in the secondary mirror as possible. It may not be perfectly cen- tered, but that is fine for now. Then, tighten the three small alignment set screws equally to secure the secondary mirror in that position.

Adjusting the Secondary Mirror’s Radial Position

Like the axial position, the secondary mirror’s radial position was set at the factory and will probably not need any adjust- ing, or if it does, you’ll typically need to do it only once.

By “radial position” we mean the position of the secondary mirror along the axis perpendicular to the focuser drawtube, as shown in Figure 39. This position is changed by adjust- ing the two spider vane thumb nuts indicated in the figure. Loosen one thumb nut, then tighten the opposite one until the secondary mirror is centered radially in the drawtube. Do not loosen the thumb nuts too much, to avoid having them

21

Image 21
Contents #9791 Orion SkyQuest IntelliScope Truss Tube DobsonianSkyQuest XX14i IntelliScope Truss Tube Dobsonian Unpacking Table ContentsNavigation Knob Box #1A IntelliScope Object Locator Kit refer to Figure Parts ListBox #3 Dobsonian Base refer to Qty. Description Box #2 Primary Mirror and Cell refer to FigureTop baseplate Assembly Assembly of the Dobsonian BaseBox #4 Truss poles refer to Figure Exploded view of the XX14i base Azimuth encoder board Brass azimuth Bushing Nylon spacer Assembly of the Optical Tube Page Ring Connector Truss poles Mounting the Optical Tube on Dobsonian Base Installing the Finder Scope and Eyepieces Adapter Thumb screws Coarse Focus Collimation Cap Aligning Collimating the Optical SystemPrimary Mirror Center Mark Preparing the Telescope for CollimatingAdjusting the Secondary Mirror’s Axial Position Adjusting the Secondary Mirror’s Radial PositionAligning the Secondary Mirror Adjusting the Secondary Mirror’s Tilt Adjusting the Secondary Mirror’s Rotational PositionStar-Testing the Telescope Aligning the Primary MirrorUsing Your Telescope Setting the Altitude TensionAltitude and Azimuth Motion Focusing the Telescope Viewing with EyeglassesAligning the Finder Scope Aiming/Pointing the Telescope Focusing the Finder ScopeMagnification 1650mmDisassembly of the Base Disassembly of the Optical TubeTube Balance Disassembling the Telescope for TransportIntelliScope Computerized Object Locator Astronomical ObservingSelecting an Observing Site Use of the Optional Light ShroudSeeing and Transparency Cooling the Telescope Let Your Eyes Dark-AdaptEyepiece Selection Moon Astronomical ObjectsCare and Maintenance Cleaning Eyepiece LensesCleaning Mirrors Page Optical Tube Assembly OTA SpecificationsAccessories MountOne-Year Limited Warranty