Kodak 6B6640 manual Filefilename

Page 25

 

 

 

Troubleshooting

 

 

 

 

Code

Status Message

Possible Cause/Subsystem

Possible Solution

 

 

 

 

1003

Unable to load default

The default configuration file was

Restore the file manually or

 

configuration file:_filename_.

deleted or renamed or moved.

reinstall the software.

 

 

 

 

1004

Unable to set values

An internal error occurred with the

Restart the application.

 

in new configuration file.

software.

 

 

 

 

 

1005

Unable to save configuration

The system is trying to save a file

Remove any unnecessary files to

 

file:_filename_.

but there is not enough disk

increase free disk space.

 

 

space.

 

 

 

 

 

1006

Configuration file field:_field_

The user entered an incorrect

• Enter the correct value.

 

is invalid.

value in a configuration file.

• Reinstall the application.

 

 

 

 

1007

Unable to remove

The user does not have the

Verify that the user has update

 

directory:_directory_.

authority to update the directory.

authority on the directory.

 

 

 

 

1101

Density DMIN/DMAX points

Density points are invalid.

Reread density values on

 

are invalid.

 

densitometer or make a new

 

 

 

calibration print.

 

 

 

 

1102

Density data is not

Density data is invalid.

Reread density values on

 

monotonic increasing.

 

densitometer or make a new

 

 

 

calibration print.

 

 

 

 

1103

Exceeded maximum number

Density data is invalid.

Reread density values on

 

of allowable data errors.

 

densitometer or make a new

 

 

 

calibration print.

 

 

 

 

1104

Asymmetric Density

Density data is invalid.

Reread density values on

 

Groupings are invalid.

 

densitometer or make a new

 

 

 

calibration print.

 

 

 

 

1201

Unknown response from

During initialization, the

Verify that the baud rate of the

 

densitometer.

application received an unknown

densitometer matches that of the

 

 

response from the densitometer.

applications configuration.

 

 

 

 

March 2001

2-9

Image 25
Contents Kodak Professional LED Page Contents Contents Calibrating the Printer LED Printer icon Starting the Device Calibration SoftwareDownloading LUTs and Sending Test Print highlights Calibrating the Printer Continue with Obtaining Densities on Obtaining Densities from a File Obtaining DensitiesUsing the Densitometer to Read Densities Open the Centralized Densitometer Application Continue with Completing the Calibration Completing the Calibration Calibrating the Printer Calibrating the Printer Calibration Graph TroubleshootingNumeric Calibration Error Codes Calibration ErrorsReadydevicename Operation Printer devicename Column, value value is Columncolumn Troubleshooting Filefilename Try adjusting frame counts or Error Message Possible Cause/Subsystem Possible Solution Non-Numeric Calibration Error MessagesCode Message Explanation Densitometer ErrorsInvalidreading Badmeastype System Requirements Installing the Calibration SoftwareMacintosh Systems Windows NT SystemsInstallation Procedure Continue Using Additional Calibration Features Application Window Definitions About the Kodak Professional Device Calibration SoftwareWindow Area Description Etc. will be deleted for this Menu Function Icon/Key Availability Description AccessMacintosh Systems NT Systems Editing the Log SettingsLog all actions Log all actions Device Setting OptionsAdding a Device Adding, Updating, and Deleting DevicesDevice icon Updating a DeviceScsi ID To delete a device and all files associated with the device Deleting a DeviceViewing Graphs Advanced Features of the LED II Calibration SoftwareValue Graph Difference Graph History Graph To view calibration graphs Applying a Master Balance NT Systems Only Using Additional Calibration Features To send a LUT to the Printer Sending LUTsUsing Additional Calibration Features To send a test target to the printer Sending TargetsNT Systems Macintosh Systems To edit the device configuration Editing the Device ConfigurationUsing Additional Calibration Features To edit the configuration information on the Procedure tab Procedure Description AttributeDensity Source Description Attribute Editing Information on the Density Source TabTo edit the configuration information on the Aim tab Only. Only use this feature as instructed by KodakAttribute Description Editing Information on the History Tab To edit the configuration information on the History tabEditing Information on the Paper Tab To edit the configuration information on the Paper tabEditing DP2 Information NT Systems To edit the configuration information on the DP2 tabEditing Kpis Information Macintosh Systems To edit the configuration information on the Kpis tab Completing the Edit Configuration File Formats Creating a Density File for Use with CalibrationExample To install the densitometer Installing the DensitometerHost System Default Port Using Additional Calibration Features Installing the Centralized Densitometer Application Installation Procedure March March Using the Centralized Densitometer Application Adding a Printer Deleting a Printer March Changing Printer Information Changing the Densitometer Port March Measuring Densities During a Calibration Cycle Viewing Calibration Graphs March Page Kodak Canada INC