Fluke 80 Series III user manual Digit Mode Model, Models 83 and 85 Bar Graph

Page 37

4-1/2 Digit Mode (Model 87)

Models 83 and 85 Bar Graph

The bar graph on Models 83 and 85 consists of 43 segments. The number of lit segments is relative to the full-scale value of the selected range. The polarity indicator at the left of the graph indicates the polarity of the input. For example, if the 40 V range is selected, the "4" on the scale represents 40 V. An input of -30 V would light the negative sign and the segments up to the "3" on the scale.

If the input equals or exceeds the 4096 counts on a manually-selected range, all segments are lit and® appears to the right of the bar graph. The graph does not operate with the capacitance or frequency counter functions.

The bar graph on Models 83 and 85 also has a zoom function, as described under "Zoom Mode".

4-1/2 Digit Mode (Model 87)

On a Model 87 meter, pressing the yellow button for one second causes the meter to enter the high-resolution,

4-1/2 digit mode. Readings are displayed at 10 times the normal resolution with a maximum display of 19,999 counts. The display is updated once per second. The

4-1/2 digit mode works in all modes except capacitance and the 250 μs and 100 ms MIN MAX modes.

To return to the 3-1/2 digit mode, press the yellow button only until all of the display segments turn on (about one second).

29

Image 37 Contents
Series Lifetime Limited Warranty Table of Contents Series List of Tables Series List of Figures Series Introduction Series International Electrical SymbolsSafety Information Terminal Your Meter’s FeaturesInputs Description Rotary Switch Positions Function Switch PositionPushbuttons Button Function Button FunctionButton Display Features Model 87 Shown Display Features Indication Number Feature100 ms ΜA, mA ΜF, nF Me, ke Hz, kHz, MHz Display Features NumberAutomatic Power-Off Power-Up OptionsMaking Measurements Input Alert FeatureMeasuring AC and DC Voltage Switch Box Testing for Continuity For in-circuit tests, turn circuit power off Measuring Resistance Isolating a Resistor DisconnectMeasuring Capacitance Using Conductance for High Resistance or Leakage TestsMeasuring Capacitance Series Testing a Diode Testing DiodesMeasuring AC or DC Current WWarningCurrent through one component Series Measuring Frequency All ranges Components of Duty Cycle Measurements Measuring Duty CycleDetermining Pulse Width Analog Bar GraphModel 87 Bar Graph Models 83 and 85 Bar Graph 2 Digit Mode ModelMIN MAX Recording Mode MIN MAX Functions Relative Mode Touch Hold ModeZoom Mode Models 83 Uses for the Zoom Mode Models 83 MaintenanceGeneral Maintenance 00.5 Ω Replace fuse OL Testing the FusesTouch top half Input contactsReplacing the Fuses Replacing the BatteryService and Parts Battery and Fuse Replacement Fluke Part Or Model Number Replacement PartsQuantity Replaceable Parts Fluke Part Number AccessoriesSpecifications KHz 20 kHz2 Models 85 and 87 AC Voltage Function SpecificationsModel 83 AC Voltage Function Specifications Model Resolution Model Model 853 Burden Voltage Typical Current Function SpecificationsΜA B Frequency Counter Specifications Capacitance and Diode Function SpecificationsAccuracy Frequency Counter Sensitivity and Trigger LevelsElectrical Characteristics of the Terminals MIN MAX Recording Specifications