Agilent Technologies e5052-90202 Calibration of the Touch Screen, Touch Panel Calibration Screen

Page 11

Option Installation

Procedure

Calibration of the Touch Screen

When you have replaced the HDD on an E5052B, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen.

Step 1.

Press

.

Step 2. Press Service Menu.

Step 3.

Press Test Menu.

Step 4.

Press Adjust Touch Screen.

 

The touch screen calibration screen (Figure 2-4) appears.

Figure 2-4

Touch Panel Calibration Screen

 

Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower

 

 

left, upper right, and lower right. Touch the x marks in that order with your finger.

 

 

Touching the four locations described above with your finger automatically concludes the

 

 

touch screen calibration.

 

 

 

NOTE

 

With no operation on the touch screen calibration screen for a preset time, it automatically

 

 

closes and the previous measurement screen reappears.

Chapter 2

11

Image 11
Contents Manufacturing No. E5052-90202 May First EditionSample bold Manual Printing HistoryGeneral Information Qty DescriptionOption Installation Introduction Installation RequirementOpen D\Recovery drive System Service Menu Administorator Menu Exit E5052B OKStep Removing Removable Hard DiskMounting Removable Hard Disk Initial Registration of E5052B Windows XP Professional Setup wizardWindows XP Professional Setup dialog box 2/2 Windows XP Professional Setup dialog box 1/2Touch Panel Calibration Screen Calibration of the Touch ScreenHard Disk Step Open D\Recovery driveRegional Sales and Support Offices