Procedure

Saving System calibration Data from Removable Hard Disk to USB Pen Drive

The following is a description of the procedures to save the system calibration data stored in the removable hard disk to a USB Pen drive.

Step 1. Exit the measurement window.

System > Service Menu > Administorator Menu > Exit E5052B > OK

Step 2. Insert a USB Pen drive into the USB port in front panel.

Step 3. Open D:\Recovery drive.

Step 4. Click .D:\syscal from the explorer.

Step 5. Select Edit - Copy from menu bar of the explorer.

Step 6. Click the USB Pen drive of the explorer.

Step 7. Select Edit - Paste from menu bar of the explorer.

Step 8. Click .D:\limit from the explorer.

Step 9. Select Edit - Copy from menu bar of the explorer.

Step 10. Click the USB Pen drive of the explorer.

Step 11. Select Edit - Paste from menu bar of the explorer.

Step 12. Turn off the E5052B, and then remove the cord from the outlet before removing the Removable Hard Disk.

Chapter 2

7

Image 7
Contents Manufacturing No. E5052-90202 May First EditionSample bold Manual Printing HistoryGeneral Information Qty DescriptionOption Installation Introduction Installation RequirementOpen D\Recovery drive System Service Menu Administorator Menu Exit E5052B OKMounting Removable Hard Disk Removing Removable Hard DiskStep Initial Registration of E5052B Windows XP Professional Setup wizardWindows XP Professional Setup dialog box 2/2 Windows XP Professional Setup dialog box 1/2Touch Panel Calibration Screen Calibration of the Touch ScreenHard Disk Step Open D\Recovery driveRegional Sales and Support Offices