Using the Configuration Page

Certain setup and configuration options are available on the passwordprotected 3M™ Attest™ Autoreader 390G Web App Configuration page. Ideally, the 3M™ Attest™ Autoreader 390G should be empty (i.e., have no incubating BIs) when it is reconfigured. Navigate to the Configuration page by entering either:

1)http://attestxxxxxx/protect/config.htm (where xxxxxx is the 6digit unit serial number), or

2)http://xxx.xx.xxx.xx/protect/config.htm (where xxx.xx.xxx.xx is the IP address)

in the web browser address bar and pressing Go or Enter. A security screen will appear. Fill in the fields as follows:

User name: manager

Password: Attest

and press OK.

Click on the desired radio buttons to change settings.

Language

Selected setting indicates language in which text on 3M™ Attest™ Autoreader 390G Web App pages will appear on the PC/mobile device

screen.

 

 

 

 

 

 

 

Selecting Enable

allows the user to configure the 3M™ Attest™ Autoreader 390G from the buttons on the unit itself.

Pushbutton Configuration

Selecting Disable

prevents the user from being able to configure the 3M™ Attest™ Autoreader 390G from the buttons on the unit

 

 

itself.

 

 

 

 

Buzzer Configuration

Selecting Enable

enables the audible alarm. The speaker icon will appear on the 3M™ Attest™ Autoreader 390G LCD panel.

Selecting Disable

disables the audible alarm. The speaker icon will disappear from the 3M™ Attest™ Autoreader 390G LCD panel.

 

 

 

Attest Autoreader 390G Nickname

Naming the 3M™ Attest™ Autoreader 390G may be useful in facilities with multiple units. The assigned nickname will appear on the

3M™ Attest™ Autoreader 390G Web App Status page.

 

 

 

Enable DHCP

Default: Setting is checked indicating the Web App will use the Dynamic IP address automatically assigned when the 3M™ Attest™

Autoreader 390G is connected to the network. Remove the check if you wish to use a Static IP address assigned by your facility IT.

 

 

 

IP Address

Enter Static IP address assigned by IT.

 

 

 

Gateway

Enter Gateway.

 

 

 

 

Subnet Mask

Enter Subnet Mask.

 

 

 

 

Primary DNS

Enter Primary DNS.

 

 

 

Use Computer Time

Check this box if you wish to set the clock onboard the 3M™ Attest™ Autoreader 390G using the computer’s clock.

 

 

Attest Autoreader 390G Time

Alternatively, leave the Use Computer Time box unchecked and complete the date and time boxes to set the

3M™ Attest™ Autoreader 390G clock.

 

 

 

 

After changing desired settings, click on Save Config button to save your changes. A new screen confirming the settings were successfully changed appears.

1

11

Image 12
Contents 3M Attest Auto-reader 390G Table of Contents 3M Attest Auto‑reader 390GContent Disclaimers WarrantySafety and Precautions PrefaceHealth and Safety Information Explanation of Product and Packaging Label SymbolsDevice Safety Compliance EMC ComplianceSpecifications Instructions for USE Initial Set‑upColor‑Coded Configuration Indicator Setting the 3M Attest Auto‑reader 390G ClockSetting the Audible Alarm Function WellsInstructions for USE ‑ Readout Procedure Instructions for USE Additional InformationInstructions for USE Positive Control Instructions for USE Interpretation of ResultsCare and Cleaning Instructions for USE DisposalTroubleshooting Guide Web Browser Requirements Serial number URLIP Address URL Dynamic IP AddressUsing the BI History ConfigurationHistory Open the Results DetailUsing the Configuration Troubleshooting the 3M Attest AUTO‑READER 390G WEB APP Repair and ReplacementMade in U.S.A. by 3M Health Care Conway Ave 3M Deutschland GmbH