3M 390G Wells, Buttons, Information Display, Color‑Coded Configuration Indicator, BI Crusher

Page 6

INSTRUCTIONS FOR USE – Wells, Buttons, & Display Interpretation

1

Please familiarize yourself with the face of the 3M™ Attest™ Autoreader 390G:

Wells

1.Incubation/Reader Wells – The 3M™ Attest™ Autoreader 390G has ten incubation wells.

Buttons

2. Information / Up Arrow Button

3. Temperature / Down Arrow Button

4. Buzzer Alarm OFF / Enter Button

Information Display

5.LCD Display – Display is backlit

6.LCD Display – Example of Error Message for well #6

7.LCD Display – Example of Remaining Incubation Time, in minutes, for well #10

8.LCD Display – Example of Negative ( ) BI result for well #9

9.LCD Display – Example of Positive ( + ) BI result for well #8

10.LCD Display – Alarm Enabled Indicator

ColorCoded Configuration Indicator

11. Incubation well configuration sticker is colorcoded green

BI Crusher

12. BI Crusher Well

INSTRUCTIONS FOR USE – 3M™ Attest™

Autoreader 390G Configuration

Note: Ideally, the 3M™ Attest™ Autoreader 390G should be empty (i.e., have no incubating BIs) when it is reconfigured.

Press the and buttons on the front of the 3M™ Attest™ Autoreader 390G simultaneously for 3 seconds to enter configuration mode. Once in configuration mode,

use the Up and Down buttons to set the 3M™ Attest™ Autoreader 390G

clock and audible alarm as desired. The button acts as an “Enter” key and moves the flashing cursor to the next changeable setting.

Setting the 3M™ Attest™ Autoreader 390G Clock

The 3M™ Attest™ Autoreader 390G date and time are displayed in this format:

YYYYMMDD HH:MM, with the time displayed using a 24 hour clock. To change the year,

if necessary, push the up or down buttons to set desired year, then press to advance

to the month, adjust as necessary and press to advance to the day. Use the button to move to the time and set the hours and minutes.

Setting the Audible Alarm Function

Move across the top row of text by pushing the button. The flashing “B” in Buzzer indicates this is a changeable setting. It is recommended this be set at “Buzzer Enabled” to audibly alert the operator to a positive BI but it may be set to “Buzzer Disabled” by using either the up or down button. Note: If Buzzer Disabled is selected, and the changes accepted, the speaker icon ( ) will disappear from the LCD panel and the alarm will be disabled for all positive 3M™ Attest™ Rapid Readout Biological Indicators, error codes, and caution code alarm warnings. Note: The 3M™ Attest™ Autoreader 390G display backlight flashes whenever an audible alarm sounds. Disabling the audible alarm does NOT disable the LCD display backlight flashing.

Press to select the desired option and advance to the Accept screen.

Accept Screen

Use the up or down buttons to toggle between Yes and No, press to accept changes (Yes selected) or retain current settings (No selected) and exit the configuration mode. The LCD panel will return to the standard display.

5

Image 6
Contents 3M Attest Auto-reader 390G Table of Contents 3M Attest Auto‑reader 390GSafety and Precautions Content DisclaimersWarranty PrefaceDevice Safety Compliance Health and Safety InformationExplanation of Product and Packaging Label Symbols EMC ComplianceSpecifications Instructions for USE Initial Set‑upSetting the Audible Alarm Function Color‑Coded Configuration IndicatorSetting the 3M Attest Auto‑reader 390G Clock WellsInstructions for USE Positive Control Instructions for USE ‑ Readout ProcedureInstructions for USE Additional Information Instructions for USE Interpretation of ResultsCare and Cleaning Instructions for USE DisposalTroubleshooting Guide IP Address URL Web Browser RequirementsSerial number URL Dynamic IP AddressHistory Using the BI HistoryConfiguration Open the Results DetailUsing the Configuration Troubleshooting the 3M Attest AUTO‑READER 390G WEB APP Repair and ReplacementMade in U.S.A. by 3M Health Care Conway Ave 3M Deutschland GmbH