Tripp Lite HCRK-1 owner manual Mounting, Battery Module Warnings, Labeling Symbols

Page 3

Important Safety Instructions continued

Battery Module Warnings

Do not allow objects to contact the Battery Module’s two DC input terminals. Do not short or bridge these terminals together. Serious personal injury or property damage could result.

It is normal for sparks to occur when connecting the Battery Module to the Power Supply Module.

Due to the possibility of dangerous arcing, do not unplug the Battery Module from the Power Supply Module while the Power Supply Module is operating from battery power (whenever the cart is mobile and the Power Supply Module is not plugged into a live utility wall outlet).

Your Battery Module does not require routine maintenance. Do not open your Battery Module for any reason. There are no user-serviceable parts inside. The batteries in the Battery Module will eventually wear out and be unable to provide adequate backup times. Internal battery replacement must be performed by factory authorized service personnel. Because batteries present a risk of electrical shock and burn from high short-circuit current, service personnel should observe proper precautions. Unplug the Battery Module before performing battery replacement. Do not short or bridge battery terminals with any object. Use tools with insulated handles. Remove metal objects such as watches and rings before working with batteries. Do not open batteries. Replace batteries only with sealed lead-acid batteries of the same number, type, voltage and amp-hour capacity. Contact Tripp Lite for details.

Tripp Lite offers additional replacement Battery Modules (model #: BP12V82). Contact Tripp Lite for details.

The Battery Module’s internal batteries are recyclable. Refer to local codes for disposal requirements, or in the USA only call 1-800-SAV-LEAD or 1-800-8-BATTERY (1-800-8-228-8379) or visit www.rbrc.com for recycling information. Do not dispose of the batteries in a fire.

If connecting a replacement Battery Module to the Power Supply Module, the Battery Module must have the same voltage as the Power Supply Module. Use only Tripp Lite Battery Modules

(model #: BP12V82) with the Power Supply Module.

Labeling Symbols

AC Voltage: V~

DC Voltage: V

Neutral Fusing:

Electric Shock Hazard:

Mounting

WARNING!

Mount the Power Supply Module and Battery Module BEFORE connection.

The mounting brackets are only designed to support a Tripp Lite Power Supply Module and a single Battery Module. Do not mount or place any other devices on the mounting brackets. The mounting brackets will not support the additional weight and may fail.

Failure to follow these warnings may lead to product damage and/or risk of personal bodily harm.

The mounting bracket is only designed to fit specific mobile medical cart configurations (see diagrams). Tripp Lite offers additional mounting brackets for additional medical cart configurations. Contact Tripp Lite for details. The user must determine the fitness of the mounting hardware and procedures before mounting. If the hardware or procedures are not suitable for your application, contact Tripp Lite for assistance.

3

Image 3
Contents Medical-Grade Mobile Power Retrofit Kit Equipment Connection Warnings Important Safety InstructionsPower Supply Module & Battery Module Location Warnings Power Supply Module Connection WarningsBattery Module Warnings Labeling SymbolsMounting Remote User Interface RUI Mounting Power Supply Module Battery Module MountingConnection / Start-Up Connect the Battery Module to the Power Supply ModulePlug In the Power Supply Module Charge the Battery ModuleTurn the Power Supply Module’s Outlets on Plug Your AC Equipment into Power Supply Module’s OutletsOPTIONAL-Install PowerAlert Data-Saving Software Stationary Operation Plugged In / Battery Charging Mobile Operation Unplugged / Battery DischargingRemote User Interface RUI Power Supply Module Battery Module Power button after reconnection. See Connection / Start-Up Storage ServiceWarranty Registration Limited Warranty