Hasselblad Flextight 646 manual White Calibration for Reflectives

Page 22

English

22

Hardware Reference - Flextight 646

Calibrating the Scanner

 

White Calibration for Reflectives

The Flextight scanner uses a white reference to make sure that all elements in the scanner's CCD react consistently to light brightnesses. The scanner has been calibrated from the factory, and the calibration data is stored in the scanner’s flash PROM.

As the scanner ages, the properties of the light tubes may change slightly. Therefore, you must redo the white calibration every six months or so and immediately after changing the light tubes.

Use procedure below to make White Calibration:

1.Find the white calibration sheet that came with your scanner. It is a square white piece of paper (219 x 219 mm (8.6" x 8.6")).

If you have lost the white calibration sheet, then you can use a flat (non-textured), perfectly white piece of paper of the same dimensions.

2.If you have used the reflective holder before, make sure that there are no finger prints or other marks on the plastic layer. Replace the layer with a new one if necessary.

(Please note that the A4/letter-size reflective holder in not included in the standard delivery).

3.Align the top edge of the sheet with the dotted line near the top of the original holder (see Figure 8 opposite). Align each side of the target with the edges of the original holder.

4.Select White Calibration from the Maintenance window. The External White Calibration window appears.

User's Guide - Flextight 646

Image 22
Contents Users Guide Users Guide Flextight Table of Contents English Hardware Reference Flextight English System Requirements Hardware Reference Flextight Users Guide Flextight Front panel features of Flextight Front PanelBack Panel & Connectors Back panel features of FlextightCut-Away View Cut-away view of FlextightEnvironmental Requirements Setting up the ScannerElectrical Requirements Installation Procedure, FireWire Interface Attaching/Detaching the Feed Table Attaching/Detaching the Feed TableMounting/Dismounting of the Cabinet Mounting/Dismounting of the CabinetOperating the Flextight 646 scanner Operating the Scanner English Reflective holder, top Original Max. resolution PPI At full scanning width ResolutionsTrue Resolutions Calibrating the ScannerSelect Focus Calibration from the Maintenance menu Use procedure below to make Focus CalibrationFocus Calibration Use procedure below to make White Calibration White Calibration for ReflectivesAlign the top edge Replacing the Light Tubes Replacing the Light Tubes Follow procedure below to replace a tubeGeneral English Cleaning the External Surface of the Scanner DisposalTechnical Specifications FCC Notice CE Declaration of Conformity Declaration of conformity