Keystone 1157 instruction manual Elliptech Control Unit

Page 5

Elliptech Control Unit

The newly designed DVS-V Elliptech Soft-Touch hand control unit makes for easy operations of the vision screener. By pressing the appropriate buttons on this compact panel, the stereo tests can be advanced or re- versed, the subject’s eyes occluded, and the peripheral vision test lamps selectively lit. The comfortable de- sign makes it suitable for either hand-held or desk-top operation.

C

BD

A

FE

A)Occlusion is controlled by pushing separate buttons for each eye.

B)Stereo tests are listed on the control key card. A lamp next to each test description signals when that tar- get is being presented.

C)Peripheral vision target lamps are lit by pressing these buttons. A signal lamp indicates which eye and angle is being tested.

D)The day/night switch allows the examiner to toggle between simulated bright “real-life” daytime illumi- nation and dim light nighttime illumination.

E)Touch these buttons and the stereo target drum either advances or reverses to the next test. Keep button depressed and drum will continue to rotate.

F)Head Sensor override button allows the operator the option of operating the screener without the use of the sensor. Overriding the sensor allows the operation of the machine without an examinee’s head prop- erly placed inside the headrest. Overriding the sensor is not recommended.

3

Image 5
Contents Keystone View Contents DVS-V Test Instrument DVS-V Maintenance Elliptech Control Unit Test Guide Test #1 Quick Acuity at 20/40 Level Test Questions and Referral GuideTest #2 Color Deficiency Quick Acuity TestDepth / Signs Test #3 & #4 Depth Perception & Sign Recognition100 200 Test #5 Binocular AcuityTest #6 Right Eye Acuity Phoria Test #7 Left Eye AcuityTest #8 Phoria 55 70 Horizontal Field Testing85 70 55 Nasal Dickerson Road, Reno NV Phone 775 324-2799 800 Fax 775