Sierra Monitor Corporation PF2A7H instruction manual Do not operate beyond specification range

Page 3

SAFETY

The Digital Flow Switch and this manual contain essential information for the protection of users and others from possible injury and property damage and to ensure correct handling.

Please check that you fully understand the definition of the following messages (signs) before going on to read the text, and always follow the instructions.

IMPORTANT MESSAGES

Read this manual and follow its instructions. Signal words such as WARNING and NOTE will be followed by important safety information that must be carefully reviewed.

WARNING

Indicates a potentially hazardous situation

which could result in death or serious

 

injury if you do not follow instructions.

 

 

NOTE

Gives you helpful information.

WARNING

Do not disassemble, remodel (including change of printed circuit board) or repair.

An injury or failure can result.

Do not operate beyond specification range.

Fire, malfunction or switch damage can result.

Do not operate in a combustible gas or explosive gas atmosphere.

Fire or an explosion can result.

This flow switch is not an explosion proof type.

Do not use with a combustible fluid.

Otherwise, a fire or an explosion or damage may potentially result. This flow switch is for air only.

NOTE

Follow the instructions given below when handling your flow switch. Otherwise, the switch may be damaged or may fail, thereby resulting in malfunction.

Do not drop it, bring it into collision with other objects or apply excessive shock (490m/s2 or more).

Do not pull the lead wire with force nor lift the main unit by holding the lead wire. (Pulling strength less than 49N)

Wiring correctly.

Do not wiring while power is on.

Do not wire with the same circuit of power line or high-voltage line. Do not use in a place in which water, oil, or a chemical is

splashes.

Install a filter and/or mist seprator on the primary side (inlet side) if foreign matter is feared to mix in a fluid.

Flush the dust in the piping with air blow before piping the switch. Do not push the setting buttons by a sharply pointed object.

Apply the power.supply when the flow rate is zero.

Start measurement by the flow switch three seconds after turning on the power.

Maintain the switch status for measurement output before setting when initializing or setting a flow rate of the flow switch. Measure after checking impacts to the equipment.

Opening and closing of flow passage by restrictor should be within max. measured flow rate value.

2

3

Image 3
Contents Digital Flow Switch Large Flow Rate Type Contents Do not use with a combustible fluid Do not operate beyond specification rangeFire, malfunction or switch damage can result ・Wiring correctly ・Do not wiring while power is onNames and Functions of Individual Parts Model Indication MethodInstallation MountingPiping connections -68, -69 Example of Internal Circuit and WiringOutput Specification -28, -29Setting OUT1 Output SpecificationsInstantaneous switch output oU10 Function Setting Integration switch output oU11Key lock setting Output method selecting mode Display selecting modeDisplay unit selecting mode Output specification selecting modeFlow rate setting mode Flow conversion modeInstantaneous switch output oU10 Clear of Integrated Value Error Display and TroubleshootingOther Functions Flow display checkSpecification Thred G Full View with Dimensions in mm