SECTION 4—CLEANING AND DISINFECTING

2.Disassemble nebulizer chamber by turning counterclockwise and separating.

3.Fill two plastic containers or bowls:

A.Washing: Hot water/detergent solution.

B.Soaking: Hot water/vinegar solution (one part vinegar to three parts water).

4.Thoroughly clean the nebulizer and mouthpiece in hot water/detergent solution.

5.Remove from the water/detergent solution and rinse with clear hot tap water.

6.Soak in hot water/vinegar solution for 30 minutes.

7.Rinse with hot tap water again and air dry thoroughly.

NOTE: If using medical disinfectant cleaners, follow manufacturer’s instructions carefully.

8.Keep the outer surface of the tubing dust‐free by wiping regularly.

The nebulizer air tubing does not have to be washed internally because only filtered air passes through the tubing.

Part No 1150755

27

Stratos™Portable Plus

Image 27
Contents StratosPortable Plus Compressor Nebulizer System Accessories Warning Table of Contents Maintenance Word Special NotesIntroduction Register Your Product Register Online at Warranty.invacare.comFeatures Typical Product Parameters Supplemental Information CompressorExtension Cords Inspection HandlingShipping Instructions UnpackingRead ALL Instructions Before Using Important SafeguardsImmediately To Reduce the Risk of ElectrocutionImportant Safeguards Important Safeguards Important Safeguards Operating Instructions SetupAdding Treatment Instructions Using Batteries Battery SpecificationsDo not dispose of batteries in fire, as they could explode Conditioning the Battery Using the Battery Charging the Battery Cleaning the Nebulizer Cleaning and DisinfectingCleaning and Disinfecting Cleaning the Mouthpiece Cleaning the CompressorChanging the Nebulizer Cleaning and Disinfecting Routine Care MaintenanceFilter Care StorageProblem Solution TroubleshootingTroubleshooting Limited Warranty Limited Warranty USA