Sony UP-DR80MD Important EMC notices for use in the medical environments, Warning, Warning

Page 3

3

Important EMC notices for use in the medical environments

• The UP-DR80MD needs special precautions regarding 

EMC and needs to be installed and put into service 

according to the EMC information provided in this 

instructions for use.

• The portable and mobile RF communications equipment 

such as cellular phones can affect the UP-DR80MD.

Warning

The use of accessories and cables other than those 

specified, with the exception of replacement parts sold by 

Sony Corporation, may result in increased emissions or 

decreased immunity of the UP-DR80MD.

Warning

If the UP-DR80MD should be used adjacent to or stacked 

with other equipment, it should be observed to verify 

normal operation in the configuration in which it will be 

used.

Guidance and manufacturer’s declaration-electromagnetic emissions

The UP-DR80MD is intended for use in the electromagnetic environment specified below.

The customer or the user of the UP-DR80MD should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic environment-guidance

RF emissions

CISPR 11

Group 1

The UP-DR80MD uses RF energy only for its internal 

function. Therefore, its RF emissions are very low and are 

not likely to cause any interference in nearby electronic 

equipment.

RF emissions

CISPR 11

Class B

The UP-DR80MD is suitable for use in all establishments, 

including domestic establishments and those directly 

connected to the public low-voltage power supply network 

that supplies buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Complies

Image 3