Guidance and manufacturer’s declaration - electromagnetic immunity

The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the 

UP-DR80MD should assure that it is used in such as environment.

Immunity test

IEC 60601 test 

level

Compliance level

Electromagnetic environment-guidance

Electrostatic 

discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or ceramic tile. If floors are 

covered with synthetic material, the relative humidity should 

be at least 30%.

Electrical fast 

transient/burst

IEC 61000-4-4

±2 kV for power 

supply lines

±1 kV for input/

output lines

±2 kV for power 

supply lines

±1 kV for input/

output lines

Mains power quality should be that of a typical commercial or 

hospital environment.

Surge

IEC 61000-4-5

±1 kV differential 

mode

±2 kV common 

mode

±1 kV differential 

mode

±2 kV common 

mode

Mains power quality should be that of a typical commercial or 

hospital environment.

Voltage dips, short 

interruptions and 

voltage variations 

on power supply 

input lines

IEC 61000-4-11

< 5% 

U

T

 

(> 95% dip in 

U

T

) 

for 0.5 cycle

40% 

U

T

 

(60% dip in 

U

T

) 

for 5 cycles

70% 

U

T

 

(30% dip in 

U

T

) 

for 25 cycles

< 5% 

U

T

 

(> 95% dip in 

U

T

) 

for 5 sec

< 5% 

U

T

 

(> 95% dip in 

U

T

) 

for 0.5 cycle

40% 

U

T

 

(60% dip in 

U

T

) 

for 5 cycles

70% 

U

T

 

(30% dip in 

U

T

) 

for 25 cycles

< 5% 

U

T

 

(> 95% dip in 

U

T

) 

for 5 sec

Mains power quality should be that of a typical commercial or 

hospital environment. If the user of the UP-DR80MD requires 

continued operation during power mains interruptions, it is 

recommended that the UP-DR80MD be powered from an 

uninterruptible power supply or a battery.

Power frequency 

(50/60 Hz) 

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at least 

characteristic of a typical location in a typical commercial or 

hospital environment.

NOTE: 

U

T

 is the a.c. mains voltage prior to application of the test level.