Guidance and manufacturer’s declaration - electromagnetic immunity

The UP-DR80MD is intended for use in the electromagnetic environment specified below. The customer or the user of the 

UP-DR80MD should assure that it is used in such as environment.

Immunity test

IEC 60601 test 

level

Compliance level

Electromagnetic environment-guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 Vrms

3 V/m

Portable and mobile RF communications equipment should 

be used no closer to any part of the UP-DR80MD, including 

cables, than the recommended separation distance 

calculated from the equation appliance to the frequency of 

the transmitter.

Recommended separation distance

d

 = 1.2 

√

P

d

 = 1.2 

√

P

   80 MHz to 800 MHz

d

 = 2.3 

√

P

   800 MHz to 2.5 GHz

Where 

P

 is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter 

manufacturer and 

d

 is the recommended separation distance 

in meters (m).

Field strengths from fixed RF transmitters, as determined by 

an electromagnetic site survey, a should be less than the 

compliance level in each frequency range. b 

Interference may occur in the vicinity of equipment marked 

with following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 

assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the UP-DR80MD is used exceeds the applicable RF 

compliance level above, the UP-DR80MD should be observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as reorienting or relocating the UP-DR80MD.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Image 5