Siemens LU01 instruction manual P924 Volume Measurement, P927 Distance Percent, Master Reset P999

Page 90

Parameters

P924 Volume Measurement

The calculated vessel capacity in Max Volume (P051) or % of Max Volume.

Press ENTER as required to display in Max. Volume units.

P927 Distance Percent

The distance between the surface and the transducer face.

Use P923 unless the distance information is required in percent (press ENTER ).

Master Reset (P999)

This feature resets all parameters to original values.

Values

Range: 0.000 to 9999

 

 

Use this feature prior to initial programming if arbitrary Parameter Values were used during a bench test, or after upgrading the software. Following a Master Reset, complete reprogramming is required.

To perform a Master Reset, access P999 and press CLEAR . C.ALL displays until the reset is complete.

CAUTION: Be careful when using this feature. All data will be reset. For convenience, be sure to record the values you want to re-enter.

Page 84

SITRANS LU01 – INSTRUCTION MANUAL

7ML19985BE02

Image 90
Contents Sitrans Manual for agreement with Technical data subject to changeThis document is available in bound version Instrumentation described, variationsTable of Contents Independent Relay Setpoints 100 Page Introduction IntroductionSmartLinx RS-232 or LocalSitrans LU01 Features Programmable FeaturesManual UnitsAmbient temperature SpecificationsSpecifications Installation categoryProgrammer OutputsCable TransducerOptions Refer to associated product documentationSafety marking symbols Manual On Product DescriptionLocation InstallationInstallation Cable/Conduit Entry RequirementsTransducer Mounting Transducer TypesM 1.65 ft Interconnection Communication Mode LEDSwitchterminal blockboard B RJ-11 port AC ModelDC Model TB1Optional SmartLinx Module LCD CardBoard a System Diagram TransducerMA Outputs Temperature SensorRelays To TS-3Level System Synchronization CommunicationSmartLinx SerialDC Supply Wiring PowerAC Supply Wiring Position. SelectUse the hand programmer to change individual parameters Communications AccessProgrammer RJ-11 port RJ-11 cableDisplay Segment ProgrammingDisplay DescriptionKeypad ProgrammingFunction keys Program Mode Entry Changing ParametersParameter Reset Features Parameters TypesProgramming Security Display in RUN Mode OperationOperation Operation Key Description BerP731 Performance Test Results Point Alarm Indicator RelaySystem Performance Evaluation Operation Level Space or DistanceOperation Helpful Hints Program modeParameter Reference Parameters1954 OFF programming permitted P000 LockSecures the Sitrans LU01 from changes Lated levelP002 Material Quick Start P001 to P007P001 Operation P003 Maximum Process SpeedSpecifies measurement units used for dimensional values P004 TransducerP005 Units P006 Empty P007 SpanSets the range level to be measured Application Parameters Volume P050 to P055 P050 Tank ShapeP051 Maximum Volume ExamplesUniversal Volume Calculation P052 Tank Dimension aP053 Tank Dimension L Dimension a as used in P050 Tank ShapeTypical volume calculations Entering a Volume BreakpointEntering a Level Breakpoint Display and Reading P060 to P062 P060 Decimal PositionP061 Convert Reading P070 Failsafe Timer Failsafe P070 to P072P062 Offset Reading Once activated, the Failsafe State initiates the following Material level reported when a Failsafe State is initiatedP071 Failsafe Material Level Selecting HI, LO, or HOLdRelays P100 to P119 P072 Failsafe Level AdvanceRelay reaction P100 Preset Applications Option Relay # Point # Standard Alarm as % of Span, P007P103 High High Alarm P101 High AlarmP102 Low Alarm P104 Low Low AlarmFunction Custom Relays P111 to P113P111 Relay Control Function Sets the control algorithm used to trip the relayIndependent Relay Setpoints Values For P111 Control Type Relay ControlRelays are energized and de-energized Action When P113 Relay OFF Setpoint Distance above and below the bound alarm setpointsP112 Relay on Setpoint P116 Dead BandP129 Relay Failsafe Relay Function P111 Preset P129Independent Relay Failsafe P129 To select an independent Relay Failsafe valueP201 mA Output Function MA Output P200 to P219P200 mA Output Range P203 mA Output Value / TransducerP210 0/4 mA Output Setpoint Independent mA Setpoints P210 and P211MA Output Limits P212 and P213 P211 20 mA Output SetpointP214 4 mA Output Trim MA Output Trim P214 to P215P213 mA Output Max Limit P215 20 mA Output TrimMA Output Failsafe P219 Record Temperatures P300 to P303P300 Temperature, Transducer Maximum Standard Data Logging P300 to P321P330 Profile Record P302 Temperature, Sensor MaximumProfile Records P330 to P337 Code DescriptionP331 Auto Record Enable P333 Auto Record IntervalAuto Record on and OFF Setpoints P334 to P337 P336 Auto Record Filling / Emptying P334 Auto Record on SetpointP335 Auto Record OFF Setpoint P337 Auto Record LOE TimeInstallation Records P340 to P342 Range Calibration P650 to P654P651 Sound Velocity Calibration Condition for use of this featureUsing Sound Velocity Calibration Offset CalibrationP653 Velocity Value altered when an Offset Calibration is performedP652 Offset Correction P654 Velocity at 20CView the transducer temperature in C Temperature Compensation P660 to P664P664 Temperature P660 Temp SourceP701 Max Empty Rate Rate P700 to P707P700 Max Fill Rate P003 Value Meters/MinuteP702 Filling Indicator P703 Emptying IndicatorP704 Rate Filter Measurement Verification P710 to P713 P705 Rate Update TimeP706 Rate Update Distance P707 Rate ValueP711 value P712 preset value P711 Echo LockP712 Echo Lock Sampling P713 Echo Lock Window P727 Scan DelayTransducer Scanning P725 to P729 P726 Level System SyncP729 Scan Time P728 Shot DelayDisplay P730 to P733 P730 Auxiliary ReadingAdjusts the Point Number display scroll speed P732 Display DelayP733 Scroll Access P731 Auxiliary Reading KeyP740 Peripheral Communications Peripheral Communication Support Parameters P740 to P749SmartLinx Reserved 750 to P748 RS-485 TerminationCommunications P772 SmartLinx Hardware TestingP790 Hardware Error P772 Baud RateEcho Processing P800 to P807 P791 Bus ErrorP792 Bus Error Count P794 SmartLinx Module Type P795 SmartLinx ProtocolP802 Transducer with Submergence Shield P803 Shot / Pulse ModeP801 Range Extension Determines what type of ultrasonic shots are firedDisplay Description P804 Confidence ThresholdP805 Echo Confidence Determines which echoes are evaluated by softwareP807 Noise Advanced Echo Processing P810 to P825P806 Echo Strength Anatomy of an Echo ProfileScrolling Display SymbolP810 Scope Displays ValuesP817 Profile Pointer Time Profile Pointer P817 to P825P816 Echo Time Raw P818 Profile Pointer DistanceP819 Profile Pointer Amplitude P820 AlgorithmP823 Reform Echo P821 Spike FilterP822 Narrow Echo Filter Dampens spikes in the echo profile to reduce false readingsP830 TVT Type Advanced TVT Adjustment P830 to P835P825 Echo Marker Trigger Selects the TVT Curve usedP831 TVT Shaper P832 TVT Shaper AdjustAllows manual adjustment of the TVT curve To change a breakpointP835 TVT Slope Min P833 TVT Start MinP834 TVT Start Duration Range -30 toP840 Short Shot Number Advanced Shot Adjustment P840 to P852Adjust the short shot transmit pulse frequency in kHz P841 Long Shot NumberP845 Long Shot Width P843 Long Shot FrequencyP844 Short Shot Width P850 Short Shot BiasP852 Short Shot Range Test P900 to P913P851 Short Shot Floor P900 Software Revision NumberP904 Keypad P902 WatchdogP903 Display P905 Transmit PulseP907 Programmer Interface P912 Transducer TemperatureP913 Sensor Temperature P910 RelayP921 Material Measurement Measurement P920 to P927P920 Reading Measurement P922 Space MeasurementThis feature resets all parameters to original values P924 Volume MeasurementMaster Reset P999 P927 Distance PercentEcho Processing Technical ReferenceTransmit Pulse Technical ReferenceEcho Processing Displays Scope Displays, P810 Distance CalculationSound Velocity Dolphin Plus DisplayVolume Calculation Volume LevelUniversal, Linear P050 = Maximum Process Speed Universal, Curved P050 =Parameter Ment P728 Shot Delay s P803Application Examples P003 ValueExample 1 Level Measurement Material LevelWhen satisfied with system performance and programming NumberInstructions Example 2 Space Measurement Material VolumeP051 Max Volume Enter 1457 for 1457 m Description InstructionsNo entry required. preset for 20 mA = low level, P001 = P001 P002 P003 P004 P005 P006 P007 P070P200 MA Range P213 MA Max Limit P219 MA Failsafe Reference Number Communication SupportApplication Assistance TechnicalMT-00 Measurement Message Technical Reference Field NameDefinition Temp Percent End Of Message # # = rate alarm = band alarmMT-01 Hold Message Start of message Message typeMessage Type Point # End Of Message MT-03 Point Not ScannedStart of message STX $02Maintenance MaintenanceUnit Repair and Excluded Liability Troubleshooting Guide Symptom Cause ActionTroubleshooting Measurement Difficulties To display Echo Confidence in the RUN modeTo update the value displayed after each aiming adjustment Flashing LOE DisplayFixed Reading Wrong Reading Troubleshooting Name Programming ChartsProgramming Charts P000 Lock GP200 MA Range P201 P116 Bound Alarm Deadband P129 Relay FailsafeP330 Profile Record P331 Auto Record Enable G P332 P202 MA Allocation P203Programming Charts Parameter Advanced Echo Processing 7ml19985BE02

LU01 specifications

The Siemens LU01 is a state-of-the-art train control and signaling system designed to enhance the safety and efficiency of rail networks. This sophisticated system leverages advanced technologies and innovative features to ensure seamless train operations, lower maintenance costs, and improved overall performance.

One of the main features of the Siemens LU01 is its implementation of the European Train Control System (ETCS). ETCS is a vital component in modernizing rail infrastructure across Europe, aiming to standardize various train control systems for enhanced interoperability among different national networks. The LU01 facilitates the integration of ETCS Level 1 and Level 2, allowing for both trackside equipment and onboard systems to work in harmony to manage train movements effectively.

Additionally, the LU01 employs a decentralized architecture that enhances system reliability and minimizes downtime. This design means that it can operate even in the case of individual component failures, ensuring that train operations remain uninterrupted. Furthermore, the LU01 utilizes advanced diagnostic tools and predictive maintenance technologies to monitor system performance continuously. This proactive approach allows for early identification of potential issues and timely maintenance, which contributes to reducing operational costs.

Another critical characteristic of the Siemens LU01 is its capability for real-time data transmission. The system integrates wireless communication technologies, enabling constant connectivity between trains and control centers. This feature allows for real-time updates on train positions, speeds, and other critical operational data, thereby enhancing situational awareness for train operators and maintaining optimal schedules.

Moreover, the LU01 is designed to be compatible with existing rail infrastructure, ensuring a smooth transition for railway operators looking to upgrade their signaling systems without extensive overhauls. This adaptability makes it an appealing choice for both new projects and retrofitting existing rail networks.

In summary, the Siemens LU01 stands out for its cutting-edge features, including support for ETCS, decentralized architecture, real-time data communication, and compatibility with legacy systems. With these technologies and characteristics, the LU01 significantly contributes to the safety, efficiency, and sustainability of modern rail transport systems, making it a pivotal solution in the evolution of railway operations.