Agilent Technologies E4400-60230, E4400-60597 manual Check Signal Generator Functionality

Page 15

Agilent Technologies Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal Generators

Check Signal Generator Functionality

Check Signal Generator Functionality

Use the following procedure to confirm that the signal generator powers up and that the internal check identifies no errors. The internal check evaluates the operation of the signal generator and returns an error message if it detects a problem.

NOTE

The signal generator must warm up for at least five minutes and all errors must be cleared from the error

 

queue before a firmware upgrade can be successfully completed.

 

If errors exist in the error queue and a firmware upgrade is attempted, the firmware upgrade will be

 

unsuccessful; new DSP code cannot be upgraded until the error queue has been cleared.

 

When signal generators with Option 1E5 are first connected to ac line power, the error message

 

ERROR 514, Reference Oven Cold occurs, causing both the OVEN COLD annunciator and the ERR

 

annunciator to turn on.

 

After approximately five minutes, the OVEN COLD annunciator automatically clears, but the ERR

 

annunciator remains on until all errors are cleared from the error queue.

 

 

To check signal generator functionality

1.Turn on power to the signal generator and let it warm up for at least five minutes.

2.Cycle power on the signal generator. The green LED should light and the signal generator should perform a functionality check.

3.When the display lights, check to see if the ERR annunciator is on.

If the ERR annunciator is on, review the error messages in the error queue by pressing

Utility > Error Info > View Next Error Message. The first error message in the error queue appears in the display text area. (Refer to the signal generator error messages list on the documentation CD for information about each error message.)

After resolving all problems causing errors, press Clear Error Queue(s).

If the ERR annunciator is off, the signal generator functionality check has passed.

15

Image 15
Contents Firmware Upgrade Guide Copyright 1999 2005 Agilent Technologies Introduction For Agilent E44xxB ESG Signal Generators Models OnlyOverview of Upgrade Assistant To check signal generator functionality Check Signal Generator FunctionalityHard Disk Space Required Confirm Required EquipmentQuantity Description Part Number Gpib Interface RequirementsConnect an RS-232, GPIB, or LAN Interface to a PC Utility GPIB/RS-232/LAN LAN is not available on E44xxB Install PSG/ESG Upgrade Assistant 5.12 and Firmware Files Use the PSG/ESG Upgrade Assistant 5.12 with Firmware Files Check Signal Generator Functionality ESG-B Firmware Installation For Versions B.03.20 Safety ConsiderationsRequired Equipment Check Signal Generator Functionality Installation Procedure Helpful Notes for Using this Procedure

E4400-60597, E4400-60230 specifications

Agilent Technologies, a renowned name in the field of measurement and instrumentation, has made significant contributions to various industries with its cutting-edge solutions. Among their highlighted offerings, the Agilent E4400-60230 and E4400-60597 models stand out due to their advanced features and technologies.

The E4400-60230 is a versatile signal generator known for its wide frequency range and exceptional signal integrity. Designed for engineers engaged in RF and microwave testing, this device offers frequency coverage from 250 kHz to 3 GHz, making it suitable for a diverse range of applications including telecommunications, automotive, and aerospace. One of its key features is the modulation capabilities, which include AM, FM, and pulse modulation, allowing users to simulate various communication scenarios effectively.

In terms of user experience, the E4400-60230 boasts an intuitive interface supported by a large display that provides clear visualization of settings and parameters. The device is also compatible with GPIB, USB, and LAN interfaces, enabling seamless integration into automated test environments. Furthermore, its low phase noise specifications ensure that users can achieve high-performance testing, particularly critical for sensitive applications.

On the other hand, the Agilent E4400-60597 is designed with an emphasis on high-frequency applications, covering a range from 250 kHz to 1.1 GHz. This model is particularly popular in the production and development of wireless communication devices. It features enhanced output power capabilities, thereby allowing users to generate cleaner and more stable signals. The E4400-60597 also supports a variety of modulation formats, ensuring that it meets the demanding requirements of modern communication standards.

Both the E4400-60230 and E4400-60597 incorporate advanced signal generation technologies that leverage digital processing to improve performance and reliability. The devices are equipped with a built-in counter and frequency sweeps for expansive testing capabilities. Additionally, their compact form factor allows for ease of portability and installation in various setups.

Together, these two models epitomize the commitment of Agilent Technologies to deliver high-quality, innovative, and versatile test equipment that meets the evolving needs of the industry. Their design and engineering excellence provide users with the tools necessary to push the boundaries of technology, ensuring precision and efficiency in their testing processes. Whether for research and development or production environments, the E4400-60230 and E4400-60597 continue to be indispensable assets for engineers worldwide.