Agilent Technologies E4400-60597, E4400-60230 Connect an RS-232, GPIB, or LAN Interface to a PC

Page 8

Agilent Technologies Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal Generators

Firmware Upgrade Kits E4400-60230 and E4400-60597 for PSG/ESG Signal Generators

Connect an RS-232, GPIB, or LAN Interface to a PC

To connect the signal generator to a PC using an RS-232 interface (Only for use with Agilent E44xxB ESG Signal Generators)

1.Connect Agilent E44xxB ESG signal generators to the PC as shown in the following diagram. Use a 5-mm nut driver to remove both standoffs (screws) from the female-female adapter.

RS-232 communications are connected to one of the COM ports on the PC using an RS-232 serial cable with a null-modem connected between the COM port and the signal generator’s RS-232 connector. Refer to the “RS-232 Installation Kit (Only for use with Agilent E44xxB ESG Signal Generators)” on page 7 for more information.

NOTE

Ensure that the selected COM port is not used by any other device while

 

PSG/ESG Upgrade Assistant 5.12 is being used.

 

 

2.Set the RS-232 parameters as follows on the Agilent E44xxB ESG signal generators: a. Press the following keys:

Preset (factory defined)

Utility > GPIB/RS-232

RS-232 Baud Rate > 19200

RS-232 Echo > Off

RS-232 Pace >

Transmit Pace > None

Receiver Pace > Xon

RTS/CTS > RTS On

8

Image 8
Contents Firmware Upgrade Guide Copyright 1999 2005 Agilent Technologies For Agilent E44xxB ESG Signal Generators Models Only IntroductionOverview of Upgrade Assistant Check Signal Generator Functionality To check signal generator functionalityConfirm Required Equipment Hard Disk Space RequiredGpib Interface Requirements Quantity Description Part NumberConnect an RS-232, GPIB, or LAN Interface to a PC Utility GPIB/RS-232/LAN LAN is not available on E44xxB Install PSG/ESG Upgrade Assistant 5.12 and Firmware Files Use the PSG/ESG Upgrade Assistant 5.12 with Firmware Files Check Signal Generator Functionality Required Equipment ESG-B Firmware Installation For Versions B.03.20Safety Considerations Check Signal Generator Functionality Installation Procedure Helpful Notes for Using this Procedure