Agilent Technologies U2781A manual Simultaneous Synchronization SSI

Page 38

3 Features and Functions

4During this identification operation, the trigger bus is used. Hence, any triggering activities on the backplane would be blocked.

5Prior to any triggering activities, you must stop the configuration activity by sending the following command:

SYSTem:IDentity {0OFF}

Do not execute the above mentioned steps when the USB modules are in the

N O T E

process of acquiring data.

You do not need to perform the above pre-configuration if you are using the Agilent Measurement Manager software. You are only required to press the “Refresh” button.

Simultaneous Synchronization (SSI)

Simultaneous Synchronization (SSI) provides synchronization between the USB modules. Figure 3- 7 illustrates an example of SSI.

SSI allows you to set one of the modules as MASTER and others as SLAVE. MASTER module sends the SSI signal to the slave modules via the backplane trigger bus. SLAVE modules will then receive the signal and begin synchronization with MASTER module.

The SCPI command below is used to set the module as master:

CONF:SSI MAST

Send the SCPI command below to set the module as SLAVE:

CONF:SSI SLAV

N O T E

Only ONE master is assigned.

 

38

U2781A User’s Guide

Image 38 Contents
Agilent U2781A USB Modular Instrument Chassis WA R N I N G Safety Symbols Regulatory MarkingsThis Guide… Declaration of Conformity Regulatory Information for Australia/New Zealand Contents Contents Getting Acquainted Introduction Applications FeaturesChecking the Contents Product Outlook General Specifications Operating EnvironmentElectrical Specifications Power Supply AC InputMechanical Specifications Physical LayoutGetting Acquainted Installation Mount Kit Installation T EModular Chassis Installation System Requirements Hardware RequirementsDAQ Driver Installation Software RequirementsInstalling the USB DAQ driver Installation Click Finish when the installation is completed Agilent Measurement Manager Software Installation Hardware VerificationInstallation Features and Functions Features and Functions SSI timing signal Functionality USB Backplane2USB backplane block diagram 3Block diagram of Trigger Bus0...7 and Trigger Out Trigger BusTrigger Out External Trigger In Star TriggerSystem Reference Clock Chassis Temperature Monitoring 5Block diagram of temperature monitoring and fan controlFan Speed Monitoring Modules Identification Identifying Modules LocationGeorgraphical Address SYSTCDES? Simultaneous Synchronization SSI 7Synchronization between modules in the chassis Features and Functions General Maintenance General Maintenance Contact us
Related manuals