5.TRANSDUCER AND TEMPERATURE COMPENSATION

The Micro-Fuel Cell has an inherent positive temperature coefficient, the effects of which have been minimized through the implementation of a thermistor compensation circuit.

Internal electronic calibration is accomplished by TAI. However, should there by any doubt, the following procedure can be used to recalibrate the electronics. Refer to Schematic.

1.Disconnect cell.

2.Move range switch to “CAL” position.

3.Adjust R2 (designated as R2 on Schematic C70246 (C70248 for 311TCXL) and designated on the A3 PCB

module assembly as R28) for 0 ± 1 mV at output of A3, pin 6.

4.Verify that the offset is the same on all ranges.

5.Re-connect cell.

6.LEAK TESTING

If a leak is suspected in the unit, DO NOT ATTEMPT TO TIGHTEN THE

DISCONNECT FITTINGS. THE FITTINGS ARE POTTED IN EPOXY AND TIGHTENING THEM WILL BREAK THE SEALS.

To check for leaks, TAI recommends one of the following procedures:

Procedure I:

1.Purge the instrument down as low as possible.

2.Place the vent line in water and disconnect the sample.

3.Next, disconnect the vent line and place the range switch on the X1000 range.

4.The unit should stay on the X1000 range if there are no leaks.

-10-

Procedure II:

1.Purge the instrument with Nitrogen at the sample port.

2.Note the reading once it has stabilized (at least 24 hrs. on the 0-2 ppm range) and increase the flow rate.

3.If the reading goes down, the unit, or the tubing to the unit, has a leak.

Image 14
Contents 311XL Series Trace Oxygen Analyzer Model 311XL Skin Contact Table of Contents 311XL Series 311XL Series Trace Oxygen Analyzer Introduction Page Page Supporting Equipment and Services OperationCalibration Procedure. Gas requirements Gas connections Zero adjustment Maintenance Page Page Transducer and Temperature Compensation Leak TestingProduct Specification Data AccuracySensitivity System Power RequirementsRecommended Spare Parts List ReproducibilityMinimum Charge is Applicable to Spare Parts Orders Drawing List Model 311XLDrawing List Model 311TCXL

ECO#02-0227 specifications

Teledyne ECO#02-0227 represents an advancement in environmental monitoring technologies, specifically designed to address the ever-evolving requirements of water quality analysis. This innovative device is equipped with cutting-edge features that enhance its functionality, providing users with reliable data for a variety of applications ranging from academic research to industrial processes.

One of the standout features of the Teledyne ECO#02-0227 is its ability to measure multiple parameters simultaneously. It offers real-time monitoring of critical water quality indicators, including dissolved oxygen, turbidity, pH levels, and chlorophyll concentrations. The simultaneous data collection capability accelerates decision-making processes, allowing for timely interventions in environmental management.

The device leverages advanced sensor technology, ensuring high accuracy and precision in measurements. The sensors are designed to minimize interference and provide consistent data across a range of conditions. This is particularly important for applications where water quality can fluctuate significantly, such as in rivers, lakes, and wastewater treatment facilities.

Another notable characteristic of the ECO#02-0227 is its rugged design, which makes it suitable for deployment in harsh environments. It features waterproof housing and is resistant to corrosion, ensuring durability and longevity even under extreme conditions. This robustness makes the device an ideal choice for field applications, where reliability is paramount.

Moreover, the Teledyne ECO#02-0227 utilizes smart connectivity options, enabling seamless data transfer and remote monitoring capabilities. Users can easily access real-time data via wireless networks, integrating the data into broader environmental monitoring systems. This connectivity facilitates collaboration among researchers and regulators, promoting more effective management of water resources.

In addition to its technical specifications, the device is user-friendly, with a straightforward interface that allows for easy navigation through settings and data retrieval. Comprehensive documentation and support resources ensure that users can maximize the device's potential without extensive training.

Overall, the Teledyne ECO#02-0227 stands out in the field of water quality monitoring. Its multi-parameter capabilities, advanced sensor technology, robust design, smart connectivity, and user-friendly interface make it an invaluable tool for environmental professionals seeking reliable and accurate data to support their initiatives. With this device, users can monitor and protect water quality more effectively, contributing to healthier ecosystems and communities.