3.1Introduction. The Model 311XL is supplied completely assembled and ready for instant use. The Micro-Fuel Cell must be installed prior use. When a sensor is first installed, the instrument should be purged for 12-16 hours with zero gas (an inert gas such as nitrogen or argon with less than 0.1 ppm oxygen). This will remove any trace levels of oxygen from the sensor or the instrument. The integral shut-off valves in the quick disconnect sample fittings, if not disturbed, will maintain this inert atmosphere within the manifold indefinitely (the equilibrium concentration of oxygen is normally less than 1000 ppm).

When the range selector is advanced from the “OFF” position, power to the instrument’s circuitry is established. The meter will constantly respond to the residual oxygen within the integral sample passages. It is impossible to achieve a “perfect” seal of the internal sample system, and what the meter is indicating is the diffusion -- consumption balance point of the internal sample system and the Micro-Fuel Cell. This “balance” point, with a properly calibrated instrument is always within the limits of the X100 range. If the reading climbs off the limits of this scale, a leak in the manifold assembly is indicated.

TO EXTEND CELL LIFE AND MINIMIZE THE TIME REQUIRED TO MAKE THE NEXT ANALYSIS, THE INSTRUMENT SHOULD ALWAYS BE PURGED WITH THE SAMPLE OR AN INERT GAS PRIOR TO BEING TAKEN OUT OF SERVICE FOR STANDBY OR STORAGE.

3.2Calibration. The inherently constant output of the cell during its useful life precludes a definitive calibration cycle. TAI feels that the interval between calibrations should be dictated by the customer’s application. If the instrument is being used to certify the oxygen content of a product for delivery, then, a calibration prior to certification would certainly be in order. If, on the other hand, the instrument is being used to monitor or guard a sample and the evidence provided by the analyzer will in themselves determine when a calibration check is in order. The sensitivity of the analyzer should be checked at two to four week intervals.

-5-

3.2.1Calibration Procedure. Gas requirements:

Zero gas: An inert gas (nitrogen or argon) with less than 0.1 PPM oxygen concentration.

Image 9
Contents 311XL Series Trace Oxygen Analyzer Model 311XL Skin Contact Table of Contents 311XL Series 311XL Series Trace Oxygen Analyzer Introduction Page Page Operation Supporting Equipment and ServicesCalibration Procedure. Gas requirements Gas connections Zero adjustment Maintenance Page Page Leak Testing Transducer and Temperature CompensationAccuracy Product Specification DataRecommended Spare Parts List System Power RequirementsSensitivity ReproducibilityDrawing List Model 311XL Minimum Charge is Applicable to Spare Parts OrdersDrawing List Model 311TCXL

ECO#02-0227 specifications

Teledyne ECO#02-0227 represents an advancement in environmental monitoring technologies, specifically designed to address the ever-evolving requirements of water quality analysis. This innovative device is equipped with cutting-edge features that enhance its functionality, providing users with reliable data for a variety of applications ranging from academic research to industrial processes.

One of the standout features of the Teledyne ECO#02-0227 is its ability to measure multiple parameters simultaneously. It offers real-time monitoring of critical water quality indicators, including dissolved oxygen, turbidity, pH levels, and chlorophyll concentrations. The simultaneous data collection capability accelerates decision-making processes, allowing for timely interventions in environmental management.

The device leverages advanced sensor technology, ensuring high accuracy and precision in measurements. The sensors are designed to minimize interference and provide consistent data across a range of conditions. This is particularly important for applications where water quality can fluctuate significantly, such as in rivers, lakes, and wastewater treatment facilities.

Another notable characteristic of the ECO#02-0227 is its rugged design, which makes it suitable for deployment in harsh environments. It features waterproof housing and is resistant to corrosion, ensuring durability and longevity even under extreme conditions. This robustness makes the device an ideal choice for field applications, where reliability is paramount.

Moreover, the Teledyne ECO#02-0227 utilizes smart connectivity options, enabling seamless data transfer and remote monitoring capabilities. Users can easily access real-time data via wireless networks, integrating the data into broader environmental monitoring systems. This connectivity facilitates collaboration among researchers and regulators, promoting more effective management of water resources.

In addition to its technical specifications, the device is user-friendly, with a straightforward interface that allows for easy navigation through settings and data retrieval. Comprehensive documentation and support resources ensure that users can maximize the device's potential without extensive training.

Overall, the Teledyne ECO#02-0227 stands out in the field of water quality monitoring. Its multi-parameter capabilities, advanced sensor technology, robust design, smart connectivity, and user-friendly interface make it an invaluable tool for environmental professionals seeking reliable and accurate data to support their initiatives. With this device, users can monitor and protect water quality more effectively, contributing to healthier ecosystems and communities.