Omega CDE-440 manual Troubleshooting, Reading above Range

Page 11

8.Troubleshooting

1)When Calibrating the equipment, the function is not complete

Verify for humidity presence at probes header or at connection box of the equipment.

Another possibility could be the equipments interconnection cable and cell being cut or in short.

2)Reading above Range

First clean the cell and try to calibrate it. If this operation is not succefull, verify the point where the cell is connected, making sure bubbles are not present and if the solution is flowing normally.

9

Image 11
Contents CDE-440 Series Conductivity Process Cells Page Conductivity Cells Instructions ManualElectrical Mechanical Description Insetion lengthmm Process Connection Body Material SS316 Technical SpecificationsElectrical Connection InstallationCDE-442 Cell Interconnection Interconnection CableTorroydal Type Insertion TypeRetractile Insertion Cell Calibration for Low Conductivity ReadingUltra Pure Water MaintenanceCleaning CalibrationReading above Range TroubleshootingPage 4506/0908