6.Installation (cont.)

Insertion Type

For a correct conductivity measurement, it is recommended to install the cell as shown on below Illustration.

The liquid flow must go thru the Sensor and exit thru the holes located at the upper portion of the sensor. See indicative arrows shown below, showing the recommended liquid flow.

Válve 2

Válve 1

Retractile Insertion Cell

For Retractable Insertion Cell installation, allow the liquid to exit thru the holes located on the upper portion of the electrode or vice-versa.

FLOW

Torroydal Type

Torroydal cell CDE-442 generally must be installed in a Transfluency Body made out of lower magnetic permeability material. Special attention must be observed regarding the sample flow being

CELL

CELL

RESERVOIR

 

PIPING

Flanged

Threaded Point

7

Image 9
Contents CDE-440 Series Conductivity Process Cells Page Conductivity Cells Instructions ManualElectrical Mechanical Description Insetion lengthmm Process Connection Body Material SS316 Technical SpecificationsElectrical Connection InstallationCDE-442 Cell Interconnection Interconnection CableInsertion Type Retractile Insertion CellTorroydal Type Cleaning MaintenanceCalibration Calibration for Low Conductivity ReadingUltra Pure WaterReading above Range TroubleshootingPage 4506/0908