Monarch 120 instruction manual Battery and Fuse Replacement

Page 5

MODEL 120 TRUE RMS MULTIMETER

6.0Battery and Fuse Replacement:

WARNING

To prevent electrical hazard or shock, turn multimeter off and disconnect test leads before removing back cover.

MODEL 120 TRUE RMS MULTIMETER

Diode Tester:

Range

Resolution

Accuracy

Test Current

Test Voltage

Overload protection

 

 

 

 

 

 

 

 

 

 

 

1mV

1% +2

<1mA

<3.5V

600Vrms

 

 

 

 

 

 

 

 

 

Frequency (Autoranging):

 

 

 

 

6.1Opening the Instrument Case When the low battery indicator shows in the display or when a fuse blows due to an overload condition, the case must be opened to replace the battery or fuse(s).

Set the Function Selector switch to the OFF position. Turn the unit over and remove the five screws from the back as shown in Figure 1.

Turn the unit face up and carefully separate the case halves as shown in Figure 2.

Figure 2

Fuse 2 15A fuse

Figure 1

Screws

Fuse 1 0.5A fuse

Battery

Range

Resolution

Accuracy

Sensitivity

Overload Protection

 

 

 

 

 

4KHz

1Hz

0.05%+1

<1Vrms

600Vrms

 

 

 

 

 

40KHz

10Hz

0.05%+1

<1Vrms

600Vrms

 

 

 

 

 

400KHz

100Hz

0.05%+1

<1Vrms

600Vrms

 

 

 

 

 

4MHz

1KHz

0.05%+1

<3Vrms

600Vrms

 

 

 

 

 

40MHZ

10KHz

0.05%+1

<10Vrms

600Vrms

 

 

 

 

 

RPM (Autoranging):

 

 

 

 

 

 

 

 

Range

Resolution

Accuracy

Sensitivity

Overload Protection

 

 

 

 

 

40K RPM

0.01K RPM

0.05%+1

<1Vrms

600Vrms

 

 

 

 

 

400K RPM

0.1K RPM

0.05%+1

<1Vrms

600Vrms

 

 

 

 

 

4M RPM

1K RPM

0.05%+1

<1Vrms

600Vrms

 

 

 

 

 

40M RPM

10K RPM

0.05%+1

<3Vrms

600Vrms

 

 

 

 

 

400M RPM

100K RPM

0.05%+1

<10Vrms

600Vrms

 

 

 

 

 

Temperature (°C/°F)

 

 

 

 

 

 

 

Range

Resolution

Accuracy

Overload protection

 

 

 

 

 

°C

1°

-50 to 0°C: 1%+4

0 to 1000°C: 1%+3

600Vrms

 

 

 

 

 

°F

1°

-58 to 32°F: 1%+8

32 to 1832°F: 1%+6

600Vrms

 

 

 

 

 

6.2Battery Replacement

This instrument uses a single 9V, NEDA 1604 or IEC6F22 or JIS006P battery. When installing a new battery, be careful to route the battery leads through the slot in the battery enclosure. Failure to do so could result in damage to the leads upon replacement of the top half of the instrument case.

6.3Fuse Replacement

WARNING

To avoid damage to the instrument or potential electrical hazard or shock, replace fuses with exact replacements. Do not use higher rated fuses in either location.

Remove and replace the defective fuse with an exact replacement.

Fuse ratings are as follows:

Fuse 1: F0.5A/600V 10 × 38mm Fast Blow

Fuse 2: F15A/600V 10 × 38mm Fast Blow

6.4Closing the Instrument Case

Ensure the Function Selector switch has not been rotated from the position when it was removed. Check to be certain that the battery leads are safely routed through the slot provided for this purpose. Replace the top half of the instrument case, turn the instrument over, and replace the five screws in the back.

Note: Do not over tighten screws. Excessive tightening could strip the threads in the plastic case.

Capacitance (Autoranging)

 

 

Range

Resolution

Accuracy

Overload protection

 

 

 

 

4nF

1PF

1.9%+20

600Vrms

 

 

 

 

40nF

10PF

1.5%+10

600Vrms

 

 

 

 

400nF

100PF

1.5%+10

600Vrms

 

 

 

 

4uF

1nF

1.5%+10

600Vrms

 

 

 

 

40uF

10nF

1.9%+10

600Vrms

 

 

 

 

400uF

100nF

1.9%+10

600Vrms

 

 

 

 

4mF

1uF

1.9%+10

600Vrms

 

 

 

 

40mF

10uF

3%+10

600Vrms

 

 

 

 

Accuracy specification:

± ([…% of reading] + [… number of least significant digits]) at

 

18°C to 28°C (64°F to 82°F) at less than 80% RH

 

True RMS accuracy for ACV and ACA are specified from 5% to

 

100% of range. Add 1% of reading for crest factor between 1.4

 

and 3 at full scale, or less than 6 at half scale.

Electromagnetic compatibility:

Vac and Aac only:

RF field = 3V/m

Total accuracy:

Specified accuracy + 2% of range

Digital display:

4000 counts

Sampling rate:

Twice per second

Analog display:

41 segments

Sampling rate:

20 times per second

Overrange indication:

Displays ‘OL’ when value exceeds range selected

Digital Output:

Bi-directional RS232. (Cable and software optional.)

Power requirement:

9 Volt battery, NEDA 1604 or JIS 006P or IEC 6F22

Battery life:

100 hours typical. Low battery indication.

Safety:

1000V CAT III, Regulation EN61010: Part 1:1993

Dimensions:

Meter only: 198×86×38mm (7.8×3.4×1.5 in)

 

With holster: 209×94×48mm (8.2×3.7×1.9 in)

Weight:

Meter only: Approx. 430g (15.2 oz)

 

With holster: Approx. 600g (21.3 oz)

9

2

Image 5
Contents True RMS Multimeter MonarchLimited Warranty Safeguards and PrecautionsDeclaration of Conformity Table of ContentsRange Resolution Accuracy SpecificationsDC Current uA and mA Autoranging Range Resolution Accuracy IntroductionBattery and Fuse Replacement Temperature Measurements Symbol Definitions and Button LocationsOperating conditions Frequency and RPM MeasurementsResistance Measurements Function Selector Rotary Switch Range ButtonDiode Tests MAX/MIN ButtonOperating Instructions