3M 490 Using the Configuration, Pushbutton, Buzzer, Attest Auto‑reader, Primary DNS, Time

Page 12

Using the Configuration Page

Certain setup and configuration options are available on the password protected 3M™ Attest™ Autoreader 490 Web App Configuration page. Ideally, the 3M™ Attest™ Autoreader 490 should be empty (i.e. have no incubating BIs) when it is reconfigured. Navigate to the Configuration page by entering either:

1)http://attestxxxxxx/protect/config.htm (where xxxxxx is the 6digit unit serial number), or

2)http://xxx.xx.xxx.xx/protect/config.htm (where xxx.xx.xxx.xx is the IP address) in the web browser address bar and pressing Go or enter. A security screen will appear. Fill in the editable fields as follows:

User name: manager

Password: Attest

and press OK.

Click on the desired radio buttons to change settings.

Autoreader Type

Select the configuration desired for wells 15/610 by clicking on the appropriate radio button.

 

 

Language

Selected setting indicates language in which text on 3M™ Attest™ Autoreader 490 Web App pages will appear on the computer/smart phone screen.

 

 

 

Pushbutton

Selecting Enable

allows the user to configure the 3M™ Attest™ Autoreader 490 from the buttons on the unit itself.

Configuration

Selecting Disable

prevents the user from being able to configure the 3M™ Attest™ Autoreader 490 from the buttons on the unit itself.

 

 

 

 

Buzzer

Selecting Enable

enables the audible alarm. The speaker icon will appear on the LCD panel.

Configuration

Selecting Disable

disables the audible alarm. The speaker icon will disappear from the 3M™ Attest™ Autoreader 490 LCD panel.

 

 

 

Attest Autoreader

Naming the 3M™ Attest™ Autoreader 490 may be useful in facilities with multiple units. The assigned nickname will appear on the 3M™ Attest™

490 Nickname

Autoreader 490 Web App Status page.

Enable DHCP

Default: setting is checked indicating the Web App will use the Dynamic IP address automatically assigned when the 3M™ Attest™ Autoreader 490 is

 

connected to the network. Remove the check if you wish to use a Static IP address assigned by your facility IT.

 

 

IP Address

Enter Static IP address assigned by IT

 

 

 

Gateway

Enter Gateway

 

 

 

 

Subnet Mask

Enter Subnet Mask

 

 

 

 

Primary DNS

Enter Primary DNS

 

 

 

Use Computer Time

Check this box if you wish to set the clock onboard the 3M™ Attest™ Autoreader 490 using the computer’s clock.

 

 

Attest Autoreader

Alternatively, leave the Use Computer Time box unchecked and complete the date and time boxes to set the 3M™ Attest™ Autoreader 490 clock.

490 Time

 

 

After changing desired settings, click on Save Config button to save your changes. A new screen confirming the settings were successfully changed appears.

1

11

Image 12
Contents 3M Attest Auto-reader Table of Contents 3M Attest Auto‑readerWarranty Safety and PrecautionsPreface Intended USEHealth and Safety Information Explanation of Product and Packaging Label SymbolsDevice Safety Compliance EMC ComplianceSpecifications Instructions for USE Initial Set‑upColor‑Coded Configuration Indicator Setting the 3M Attest Auto‑reader 490 ClockSetting the Audible Alarm Function WellsInstructions for USE ‑ Readout Procedure Setting the Incubation Well ConfigurationBI Placement Instructions for USE Additional InformationInstructions for USE Positive Control Instructions for USE Disposal Care and Cleaning Troubleshooting GuideWeb Browser Requirements Serial number URLIP Address URL Dynamic IP AddressUsing the BI History ConfigurationHistory Open the Results DetailUsing the Configuration Nickname Auto‑reader 490 Web App Status Enable DhcpPushbutton BuzzerTroubleshooting the 3M Attest AUTO‑READER 490 WEB APP Repair and ReplacementMade in the U.S.A. by 3M Health Care Conway Ave