Ultra Sensitivity Model M200EU NO/NO2/NOx Analyzer | Addendum to M200E |
2.SPECIFICATIONS, AGENCY APPROVALS, AND WARRANTY
Min/Max Range (Physical Analog Output)
Measurement Units
Noise at Zero1
Noise at Span1
Lower Detectable Limit1
Zero Drift2
Span Drift2
Linearity
Precision
Lag Time1
Rise / Fall Time1, 3
Sample Flow Rate
Temperature Range
Humidity
Hydrocarbon Interference
Voltage Coefficient
Dimensions H x W x D
Weight, Analyzer
Weight, Ext Pump Pack
Power, Analyzer
Power, Analyzer4
Power, Ext Pump
Power, Ext Pump4
Analog Outputs
Analog Output Ranges
Analog Output Resolution
Status Outputs
Control Inputs
Min:
ppb, ug/m3 (user selectable Single, independent or auto ranges) <25 ppt RMS
<0.25% RMS of reading above 50 ppb 50 ppt RMS
<0.1 ppb / 24 hours <0.2 ppb / 7 days
<0.5% of reading or 50 ppt RMS whichever is greater / 24 hrs <1% of reading or 100 ppt RMS whichever is greater / 7 days
1% of full scale or 0.1 ppb whichever is greater 0.5% of reading above 50 ppb
20 sec
95% in <50 sec
1000 cc/min. ± 10%
Propylene rejection ratio > 20,000:1 Ethylene rejection ratio > 40,000:1
< 0.1% per V
7" x 17" x 23.5" (178mm x 432 mm x 597 mm)
18 kg (40 lbs)
9.5 kg (21 lbs)
100V~ 50/60 Hz, 120V~ 60 Hz, 230V~ 50 Hz, 2.5A, 125 watts 230V~ 50 Hz, 2.5A
110V~ 60 Hz, 220V~ 50 Hz, 240V~ 50Hz, 150 watts 230V~ 50 Hz, 2.2A
Four (4) Outputs
All outputs: 100 mV, 1, 5, 10V
Three outputs convertible to
1 part in 4096 of selected
8 Status Outputs from
2 | TAPI 05385 Rev B |
PRINTED DOCUMENTS ARE UNCONTROLLED