Hitachi vl-21a instruction manual Low Luminance Sensitivity

Page 22

9.2.4. Horizontal Resolution

TEST CONDITIONS

Refer to “Measurement Conditions”

 

 

MEASURING SYSTEM

System 1

 

 

PROCEDURE:

 

 

 

1.Shoot the resolution chart and zoom WIDE or TELE to fit the chart to the monitor.

2.Adjust the brightness and contrast of the B/W monitor so that each step of the gray scale portion of the chart can be observed.

3.Change the scan size of the monitor to under scan.

4.The reference arrows on the resolution chart are positioned at the edge of the under scanned picture.

5.Change the scan size of the monitor from under scan to over scan.

6.Measure the maximum horizontal resolution on the picture.

SPECIFICATION:

More than 480 TV Lines (High Resolution)

9.2.5. Low Luminance Sensitivity

TEST CONDITIONS

Refer to “Measurement Conditions”

 

 

MEASURING SYSTEM

System 1

 

 

PROCEDURE:

 

 

 

1.Shoot the gray scale chart and zoom WIDE or TELE to fit the chart to the monitor.

2.Adjust the brightness of the light source by using the dimmer switch so that the white

peak level of the chart becomes 30 IRE on the waveform monitor.

2. Measure the level of illumination by using the illumination meter.

SPECIFICATION:

22

Image 22
Contents Hitachi Table of Contents CN# NtscOff / On Night Mode 0.25 lux, Digital Slow ShutterColor Red Blue Yellow Burst Environment Condition and Test Measurement SpecificationsInterface Lens Test Condition Appearance Appendix Page Focus Focus Mode Manual Focal DistMenu Zoom StartExposure BLC White BAL 3D-DNR Special General Initial Exit Zoom SpeedFocus ExposurePush Focus Exposure BLC3D-DNR Special General Initial Exit ATW REDSpecial Special General Initial ExitMotion Focus Exposure BLC White BAL 3D-DNR Special Zone Select Center Zone State OFF Sensitivity Initial ExitGeneral Initial Exit ID Display OFF CAM Title Language ENG Protocol DEF Baudrate CAM IDGeneral VersionSignal System NtscCommunication Protocol Page Measurement Conditions Standard Measurement Conditions Test Conditions ProcedureSpecification Measuring SystemColor RED Blue Yellow Burst Low Luminance Sensitivity Contact Information