Ikelite 6146.10, 4103.51TTL instruction manual Lubricants, Control Maintenance

Page 4

NOTE: DS Substrobe Update information

 

DS50 Substrobes

 

 

• DS50 SubStrobes with a Serial Number below 63,850 can not be

updated to operate with the Conversion Circuitry.

 

• DS50 SubStrobes with a Serial Number between 63,850 and

69,999 operate with the Conversion Circuitry, but require an

update to provide optimum performance.

 

• DS50 Substrobes with a Serial Number of 70,000 or higher, or

with one of the two following

 

 

labels in the battery

 

ikelite

compartment provide

 

Substrobe

optimum performance with

serial number

MOD-NC

the Conversion Circuitry and

 

do not need an update.

 

 

DS125 Substrobes

 

 

• DS125 Substrobes with a Serial Number below 2,500 must be

updated to operate correctly with the Conversion Circuitry.

• DS125 Substrobes with a Serial Number between 2,501 and

4,900 will operate with the Conversion Circuitry, but require an

update to provide optimum performance.

 

• DS125 Substrobes with a Serial Number of 5,000 or higher, or

with one of the two

 

 

following labels in the

 

ikelite

battery compartment,

 

Substrobe

provide optimum

serial number

MOD-NC

performance with the

 

Conversion Circuitry and do

 

 

not need an update.

 

 

To Update Your Substrobe:

Send to the Ikelite address on the back page of this manual or contact Ikelite for more information.

13

Lubricants

1. Ikelite provides silicone lubricant with thehousing. We recommend you use only Ikelite lubricant on Ikelite products as some other brands may cause the o-ring to swel and not seal properly.

2. Use only enough lubricant to lightly cover control shafts and o-rings. Wipe off any excess lubricant with a clean cloth. Lubricant is not a sealant, it is used to reduce friction. Excessive lubricant can collect sand and dirt which may interfere with proper sealing.

CAUTION

Never use spray lubricants as the propellant ingredient can cause the plastic housing to crack.

Control Maintenance

Ikelite controls are designed to provide years of reliable service with minimal maintenance.

1. Push button controls require no maintenance other than rinsing in fresh water after saltwater use. If a push button control becomes difficult to push or if it sticks when depressed, soak the

Maintenance

The Ikelite Digital Housing should be given the same care and attention as your other photographic equipment. In addition to normal maintenance, we recommend that the housing be returned to Ikelite periodically to be checked and pressure tested.

1. Do Not leave the camera and housing in direct sunlight for prolonged periods. Heat may damage the camera.

2. Do Not ship the camera in the housing.

3. Before using the housing, always check the tightness of the set screw in each control knob.

Check each control gland penetrating the housing to make sure they are tight. There is a slight chance that either could vibrate loose during travel.

4. Keep the back o-ring clean and lightly lubricated. To lubricate, remove the o-ring from the back. Put a small amount of lkelite lubricant on your fingers. Pull the o-ring through your fingers to apply a light coating of lubricant. Only apply enough lubricant to make the o-ring feel slick. Do Not stretch the o-ring. This light coating of lubricant will help to keep the o-ring from drying out and will help to show a dark sealing line when the housing back is properly sealed.

5. Keep the area where the o-ring fits and the sealing surface of the housing clean.

6. Rinse the housing exterior thoroughly in fresh water after each salt water use. Dry with a soft cloth. Dry lens port to eliminate water spotting.

After several uses in salt water, soak the housing exterior in a mild soap solution, rinse and dry, before storing. When storing the housing, remove the back o-ring, lightly lubricate, and place in a plastic bag. Place the plastic bag with o-ring inside the housing for safe keeping.

7. If removing a housing push button, Do Not re-use the E-clip. Contact Ikelite for replacement E-clips (part #0319.12).

14

Control Maintenance Cont.

3. Some of the controls have a short shaft and cannot be pulled out exposing the shaft for lubrication. In the unlikely event one of these controls sticks or becomes difficult to operate you can remove the control from the housing and lubricate it, or return the housing to Ikelite for maintenance. To remove the control (Diagram A), loosen the set screw in the knob (allen wrench required); remove the knob. If there is salt or dirt build-up on the exposed control shaft, clean the shaft. Open the housing and gently slide the control shaft out of the control gland. Clean and lightly lubricate the shaft, including the end of the shaft. Slide the shaft back into the control gland and gently slide it back and forth a few times without fully removing the shaft from the gland. Replace the knob noting the flat area on the shaft. The set screw in the knob should tighten down against the flat area on the control, so the knob does not turn on the shaft.

housing in luke warm fresh water. After a few minutes, operate the push button. If this does not correct the problem, return the housing to Ikelite for maintenance.

2. Some of the controls have long shafts. These controls can be pulled out, exposing the shaft (see diagrams A and B on pages #16 & #17).

To lubricate the control, gently pull on the knob until the stainless steel shaft is exposed. Lightly lubricate the shaft, then

housing

control

gland

move the shaft in and out several times. This will lubricate the x-ring in the Ikelite control gland. This should be done before using the housing after a prolonged storage period, or once a week when the housing is in constant use.

15

Tighten set screw down

 

 

against this flat area

Lubricate end of shaft

Loosen set screw

when replacing the knob.

before reinserting into

 

gland

 

Diagram16 A

Image 4
Contents Ikelite Digital Housing Please ReadInstalling the Camera Closing the HousingChecking Controls Turn Camera OnUsing Flash Maintenance Control MaintenanceLubricants Photo Tips

4103.51TTL, 6146.10 specifications

The Ikelite 6146.10 and 4103.51 TTL (Through-The-Lens) systems represent cutting-edge underwater photography technology, catering to both professionals and enthusiasts. These systems are designed specifically for underwater environments, providing photographers with the tools they need to capture stunning images with depth, clarity, and precision.

A standout feature of the Ikelite 6146.10 housing is its compatibility with various DSLR cameras, ensuring that users have the flexibility to utilize their preferred equipment. This housing is constructed from robust materials, providing exceptional durability while maintaining a lightweight profile. The transparent back and ergonomic design enhance usability, allowing photographers to focus on capturing images rather than struggling with cumbersome gear.

One of the most significant advancements in these systems is the TTL functionality, which allows for automatic adjustment of the flash output. This feature ensures perfectly exposed images regardless of varying lighting conditions, making it particularly useful for underwater photography where natural light can be unpredictable. The TTL capability works seamlessly with Ikelite strobes, streamlining the process of achieving optimal exposure.

Another key aspect of the Ikelite 6146.10 is its user-friendly controls. The design incorporates easily accessible buttons and knobs, allowing photographers to make adjustments quickly even while wearing thick gloves. This is crucial in the underwater environment, where a single moment can dictate the success of a shot.

The 4103.51 TTL flash is designed to integrate flawlessly with the housing, offering robust lighting options tailored for underwater photography. It features a powerful guide number, ensuring ample illumination even in deeper waters. The strobe also offers multiple power settings and customizable flash durations, giving photographers the ability to tailor their lighting according to specific scenarios.

Waterproof to considerable depths, both the housing and strobe are engineered to withstand the rigors of the underwater environment. The use of corrosion-resistant materials prevents deterioration and ensures longevity. With precision O-ring seals, users can have confidence in the watertight integrity of the system, allowing for worry-free diving and shooting experiences.

In summary, the Ikelite 6146.10 and 4103.51 TTL systems are exemplary tools for underwater photography, featuring durable construction, TTL functionality, user-friendly controls, and robust lighting capabilities. These systems empower photographers to unlock their creativity beneath the waves, capturing vibrant images that showcase the beauty of the underwater world.