Drive Medical Design 14027 manual CPR Function, Pressure SET UP, LOW Pressure Warning

Page 12

OPERATION

!Always read the operating instructions before use.

CPR FUNCTION

When there is an emergency to perform CPR on the patient, pull the CPR strap at the head section of the mattress to release the air quickly from the mattress.

The CPR strap is located at the front right-hand side of the mattress.

PRESSURE SET UP

Users can adjust the pressure level of the air mattress to a desired firmness by themselves or according to the suggestion from a health care professional.

NOTE! It is recommended that the pressure-adjust knob turned to Firm or press Auto Firm on the panel when the mattress is first inflated. Users can then easily adjust the air mattress to a desired firmness according to the patient’s weight and comfort.

LOW PRESSURE WARNING

When abnormal pressure occurs, the Low Pressure indicator will come on. For Q2-02 and Q2-03, the audible alarm will be activated to bring attention to a low pressure condition.

Check if the connections are secure and correctly installed according to the relevant instructions.

NOTE! If the pressure is consistently low, check for any leakage (in tubes or connecting hoses). If necessary, contact your local dealer to replace any damaged tubes or hoses.

12

Image 12
Contents Item # Table of Contents Important Safeguards Read ALL Instructions Before UsingStatements & Symbols OthersIntroduction GeneralIntended USE ContraindicationPackage Content PumpMattress Cover SheetProduct Functions Installation StepGrounding Instructions Operation For productsFor patients Hand checkAlways read the operating instructions before use Pressure SET UP CPR FunctionLOW Pressure Warning Cleaning Handling & Storage LOW PressureTroubleshooting Specifications SystemWarranty Page Page Seaview Boulevard Port Washington, NY Phone Fax