2. DISASSEMBLY
NOTE FOR REPAIR
•Make sure that the flat cable and flexible board are not cracked of bent at the terminal. Do not insert the cable insufficiently nor crookedly.
•When remove a connector, don’t pull at wire of connector. It is possible that a wire is snapped.
•When installing a connector, don’t press down at wire of connector. It is possible that a wire is snapped.
Cut and remove the part of gilt which comes off at the point. (Be careful or some
pieces of gilt may be left inside)
DISCHARGING OF THE ST-185 BOARD’S CHARGING CAPACITOR (C5006)
The charging capacitor (C5006) of the
There is a danger of electric shock by this high voltage when the capacitor is handled by hand. The electric shock is caused by the charged voltage which is kept without discharging when the main power of the unit is simply turned off. Therefore, the remaining voltage must be discharged as described below.
Note:
Discharging the Capacitor
Preparing the Short Jig
To preparing the short jig, a small clip is attached to each end of a resistor of 1 kΩ /1 W
Wrap insulating tape fully around the leads of the resistor to prevent electrical shock.
1 kΩ/1 W
Wrap insulating tape.
R:1 kΩ/1 W
(Part code:
C5006
NOTE FOR DISCONNECTING THE HARNESS (COAXIAL CABLE)
When disconnecting the harness (COAXIAL CABLE), do not pull the harness part but pull off the connector body with tweezers etc.
Harness (coaxial cable)
Tweezers etc.