Emerson 54e pH, PN 51A-54epH Panel Mounting Wall Mounting Pipe Mounting, Locating the Controller

Page 4

MODEL 54e pH/ORP

WARNING

All electrical installation must conform to the National Electrical Code, all state and local codes, and all plant codes and standards for electrical equipment. All electri- cal installations must be supervised by a qualified and responsible plant electrician.

LOCATING THE CONTROLLER

Position the Model 54e pH/ORP controller to minimize the effects of temperature extremes and to avoid vibration and shock. Locate the controller away from your chemical process to protect it from moisture and fumes.

Select an installation site that is more than 2 ft from high voltage conduit, has easy access for operating personnel, and is not exposed to direct sunlight.

UNPACKING AND INSPECTION

Inspect the exterior of the shipping container for any damage. Open the container and inspect the controller and related hard- ware for missing or damaged parts.

If there is evidence of damage, notify the carrier immediately. If parts are missing, contact Rosemount Analytical customer sup- port.

MECHANICAL INSTALLATION

Mounting the Controller

The Model 54e pH/ORP controller may be supplied with a mount- ing bracket accessory. If you use the mounting bracket on wall or pipe installations, avoid mounting on pipes which vibrate or are close to the process. The bracket may be modified to mount the controller on I-beams or other rigid members. You can also fabri- cate your own bracket or panel mount the controller using the bracket as an example.

Wall or Surface Mounting:

1.Mount the bracket to the controller using the supplied four screws.

2.Mount controller mounting bracket to wall using any appro- priate fastener such as screws, bolts, etc

INSTALLATION

Pipe Mounting:

1.Attach the mounting bracket to the rear of the controller and tighten the four screws.

2.Place supplied U bolts around the mounting pipe and through the pipe mounting bracket and mounting bracket. Tighten the U bolt nuts until the controller is securely mount- ed to the pipe.

Panel Mounting:

The controller is designed to fit into a 5.43 x 5.43 inch (DIN stan- dard 144x144 mm) panel cutout (shown below). Installation requires both front and rear access.

1.Install the controller as shown below. Insert the instrument enclosure through the front of the panel cutout and align the panel mounting brackets as shown.

2.Insert two mounting bracket screws through each of the two mounting brackets and into the tapped holes in the rear of the controller enclosure and tighten each screw.

3.Insert four panel mounting screws through each hole in the mounting brackets. Tighten each screw until the mounting bracket holds controller firmly in place. To avoid damaging the controller mounting brackets, do not use excessive force.

PANEL MOUNTING

4

WALL MOUNTING

PIPE MOUNTING

Image 4
Contents Read this page Before Proceeding Risk of Electrical ShockPage Instrument Specifications @ 25C Physical Specifications GeneralInstrument Specifications @ 25C ORP Panel Mounting Wall Mounting Pipe Mounting Locating the ControllerMenus Measurement Type Output SetpointsAutocal SINGLE-POINT StandardizationSensor Wiring Diagram Power Input and Relay Output Wiring for Model 54e pH/ORP