MODEL 54e pH/ORP
PHYSICAL SPECIFICATIONS – GENERAL
Enclosure:
144 X 144 X 131 mm, DIN size (5.7 X 5.7 X 5.2 in.) Front Panel: Membrane keyboard with tactile
feedback and user selectable security. Light gray, blue and white overlay
Light gray enclosure, dark gray bezel
Display:
The display contrast is compensated for ambient temperature.
Process Variable Character Height: 16mm (0.6 in.)
Electrical Classification:
Class I, Division 2, Groups A, B, C, & D. T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.
Max. relay contact rating: 28 Vdc; 110 Vac; 230 Vac; 6 amps resistive
FM: Max. relay contact rating: 28 Vdc resistive 150 mA - Groups A & B;
400 mA - Group C;
540 mA - Group D
Power:
Code
Code
Current Outputs: Output 1: pH, ORP, temperature, glass impedance, or reference impedance.
Output 2: pH, ORP, temperature, glass impedance, or reference impedance.
Each output is galvanically isolated,
RFI/EMI: EN-61326
LVD (Code -01 only): EN-61010-1
Ambient Temperature: 0 to 50°C (32 to 122°F) NOTE: The analyzer is operable from
Relative Humidity: 95%, non condensing
Alarms:
Relay 1 - Process, Interval, or Time Proportional Control (TPC requires code
Relay 2 - Process, Interval, or Time Proportional Control (TPC requires code
Relay 3 - Process, Interval, or Time Proportional Control (TPC requires code
Relay 4 - Sensor/analyzer and process fault alarm
Each relay has a dedicated LED on the front panel.
SPECIFICATIONS
Relay Contacts: Relays
Relay 4: Epoxy sealed form C, SPDT
| Resistive | Inductive |
28 Vdc | 5.0 Amps | 3.0 Amps |
115 Vac | 5.0 Amps | 3.0 Amps |
230 Vac | 5.0 Amps | 1.5 Amps |
Weight/Shipping Weight: 1.8 kg/2.3 kg (4 lb/5 lb)
INSTRUMENT SPECIFICATIONS @ 25°C
— pH
Measurement Range: 0 to 14 pH
Output Scale Expansion: Zero suppression: up to 13 pH units
Span: Any pH from 1 to 14
Accuracy: ± 0.01 pH
Repeatability: ± 0.01 pH
Stability: ± 0.01 pH/month,
Temperature Coefficient: Input: ± 0.003 pH/°C
Output: ± 0.006 pH/°C
Temperature Compensation: Pt 100 or Pt 1000 RTD, Automatic or Manual;
INSTRUMENT SPECIFICATIONS @ 25°C
— ORP
Measurement Range:
Output Scale Expansion: Zero suppression: up to ±1300 mV
Span: Any ORP range from 100 to 2800 mV Accuracy: ± 1.0 mV
Repeatability: ± 1.0 mV
Stability: ± 1.0 mV/month,
Temperature Coefficient: Input: ± 0.2 mV/°C
Output: ± 0.4 mV/°C
Temperature Measurement:
3