3M 3MTM Volatile Organic Compounds VOCs, PID Zero Calibration, PID Span Calibration, Logging VOCs

Page 5

Volatile Organic Compounds (VOCs)

VOCs, air pollution with carbon-containing chemicals, are measured using a Photoionization detector 10.6eV. The EVM is equipped to notify you when maintenance and cleaning is required for the PID sensor. A PID sensor error will appear indicating:

(1) if the sensor needs to be cleaned, or (2) the lamp needs replacing.

PID Zero Calibration

The PID ppm zero calibration preferred method is to flush with Zero Grade Air. Alternatively, Nitrogen may be used. For best response time, it is recommended to set the flow rate at 1 LPM. NOTE: Please refer to the EVM series manual for more information on the PID ppm/ ppb sensors.

1.Connect the Zero Grade Air, tubing, and cal. cup to the EVM. (Please skip this step if you are not applying a gas.)

2.Select Calibration from the Start screen and select PID (using up/down arrows).

3.Press Cal softkey and then press Zero softkey. Once level is stabilized, press Set softkey. (Typically takes 60 seconds to stabilize.)

4.Once level has been set, you can press Save or Cancel. (Disconnect components and return to the start screen via On/Off/Esc key.)

PID Span Calibration

Typically, Isobutylene is used when the VOCs are unknown. NOTE: in order to measure

specific VOCs, you will want to use the relevant VOC for your calibrated gas source for a

ppm/ppb sensors. (Please refer to the manual for PID correction factor table; Appendix C for details.)

1.Connect Isobutylene, regulator, tubing and cal cap to the EVM.

2.Follow steps 2-4 above (in PID Zero Cal. section) but select Span softkey instep 3.

Logging VOCs

1. From the start screen or any measurement screen, press key.

To view VOCs measurements, select either the Gas or Composite screen by pressing

keys

 

from the start screen.

 

2. Press key to stop logging. (See “Viewing logged results” to access past sessions.)

Run icon

PID sensor measurement

Gas measurement screen

Image 5
Contents Quick Start Guide Start Screen and Measurement Views EVM and User InterfaceTurning On/Off and Navigating Particulate Measurements Reset/Clear memory, if necessary Start\File System menuOverview of running measurements Logging Particulates Particulate Zero CalibrationParticulate Advanced Calibration Gravimetric PID Zero Calibration Volatile Organic Compounds VOCsPID Span Calibration Logging VOCsCarbon Dioxide Gas Toxic GasesToxic and Carbon Dioxide Zero Calibration Type of Gas Range Res Zero cal Span calLogging Toxic Gases or Carbon Dioxide Measuring particulatesToxic and Carbon Dioxide Span Calibration Logging/Viewing Temperature and/or Relative Humidity Temperature and Relative HumidityNavigate to the Setup\Calibration\Flow RT rate screen Particulate Air FlowNavigate to Start \Past Session screen Air VelocityViewing Logged sessions past sessions Thin-layer of silicone is applied to a clean impactor Maintenance/CleaningSensors Locations Impactor Care and ParticulatesViewing your data Communicating074-301, Rev.D /13