Care & Maintenance
Display Unit Care
Clean with a soft, damp cloth. Do not use caustic cleaners or abrasives. Keep away from dust, dirt and moisture. Clean ventilation ports regularly with a gentle puff of air.
5-in-1 Sensor Care
Clean The Sensor
Clean with a soft damp cloth. Do not use caustic cleaners or abrasives that will mar the polished surfaces of the rain collection funnel or the solar cell. Scratches will result in decreased performance and reliability.
Insect Prevention
Insects may cause obstructions and interrupt data by nesting in or on the
Snow & Freezing Weather
The
Clean The Rain Collector Cup
Remove and empty rain collector debris filter. The debris filter is located in the rain collector funnel. Remove from the top by gently squeezing and pulling out.
Clean the Wind Vane & Anemometer
Remove foreign matter from the outside of the sensor for free movement of the wind vane and anemometer. If needed, use a small amount of spray lubricant, clear silicone or graphite powder on the anemometer for improved movement.
Calibration
Calibrate Temperature & Humidity
The indoor / outdoor temperature and humidity readings can be calibrated on the display unit to improve accuracy. Calibration improves accuracy when
1.Choose the weather category you wish to calibrate by pressing the up or down arrow buttons until the category is displayed on the Selectable Category part of the display unit.
2.Press AND HOLD the up and down arrow buttons and the “” button all at the same time for
•The display unit beeps, and all display data is hidden except for the value being calibrated.
•“CAL” (calibrate) and arrows appear next to the value being calibrated.
3.Press the up or down arrow buttons to calibrate the data value higher or lower from the actual reading.
4. Press the “ ” button to confirm the calibration adjustment. Note: The “CAL” icon remains illuminated next to calibrated values.
Calibrate the Rain Gauge
The rain gauge on the
1.First, ensure
2.Place display unit close so you can monitor it during calibration.
3.Make a pin hole in the bottom of a plastic cup. Hold the cup over the rain gauge and fill it with exactly 1 cup (8oz.) of water, allowing the water to drip into the rain gauge. You should hear the internal buckets tip and see water drain through the rain gauge.
4.A few seconds after each bucket tip, the display unit displays rainfall in approximately 0.01" or more increments.
5.The cup of water should take more than 10 minutes to empty; a quicker period will result in inaccurate calibration. Try to simulate a normal steady rainfall. When cup is empty of water, display unit should register 1.06"
Tips
•There should be nearly an equal number of water drops (about 25 water drops) between bucket tips. If not, adjust the calibration screws on the bottom of the
•If you don't hear the buckets tipping and see water dripping alternately out of each drain, there may be an issue with the rain gauge or it's adjustment. See Troubleshooting on page 15.
Adjustment
If the rain gauge doesn't register close to 1.06", make an EQUAL adjustment to the two calibration screws on the bottom of the
•To adjust the rainfall reading by 2% turn both screws 1/8 of a turn.
•To adjust the rainfall reading by 4% turn both screws 1/4 of a turn.
•To adjust the rainfall reading by 8% turn both screws 1/2 of a turn.
16 | 17 |