Omega Speaker Systems OME-DB-889D manual Calibration VR Description, Calibration Steps

Page 25

6. Calibration

The OME-DB-889D is calibrated to its best of operation. For environment subjected to large vibrations, recalibration is recommended. Before calibrating the OME-DB-889D, user should consider having the following available:

zOne 6 digital voltage meter

zOne temperature meter

zOne OME-A-82x series multifunction card

6.1 Calibration VR Description

There are two VRs on the OME-DB-889D.

VR Num.

Description

VR1

Amplifier Offset adjustment

VR2

CJC Gain adjustment

6.2 Calibration Steps

1.

2.

3.

4.

5.

6.

7.

8.

9.

10.

11.

Turn off computer power.

Connect OME-A-822PGL (Or any OME-A-82x series multi-function card) to OME-DB-889D.

Setup OME-DB-889D to default setting

Short CH0HI to CH0LO to A.GND of DB889D.

Connect probe of voltage meter to CN7 (CJC output: + , A.GND : - ). Turn on computer power

Wait about 5 minute (warn up).

Run Calibration program of OME-DB-889D.

Close the probe of temperature meter to OME-DB-889D them reading temperature value.

Adjust VR1 until the screen value = 0;

Adjust VR2 until voltage meter reading value = temperature value X 2.44mV

OME-DB-889D User’s Manual

23

Image 25
Contents User’sGuide Servicing North America OME-DB-889D User’s Manual Specification FeaturesOME-DB-889D Multiplexer Board Function Description ApplicationsFSR Layout Installing1 JP0 ~ JP15 Analog Input R/C Filter Jumper setup3 JP19 Internal / External Power selection Local Gain Control Sw1 Pin Assignment Page Input Channel Selection Analog input & Gain settingGain Setting Gain = OME-A-82X Gain X OME-DB-889D Gain Thermocouple InputVoltage to Temperature Conversion TC = V*1000/24.4 CJC OutputFloating Signal Connection Signal ConnectionNon-Floating Signal Source OME-A-82X OME-DB-889D Cascading OME-DB-889DOpen Thermocouple Detection Block Diagram Programming Using OME-A-822PGLDEMO01.C Example ProgramPage Calibration VR Description CalibrationCalibration Steps WARRANTY/DISCLAIMER Temperature