Teledyne TED 191 instruction manual Section SIX Troubleshooting

Page 18

SECTION SIX

 

TROUBLESHOOTING

 

 

 

SYMPTOM

 

CURE / REMARK

 

 

 

 

1)

Check battery drawer is properly located.

No numeric display on

2)

Check battery polarity .

 

 

the LCD when powered

3)

Check batteries for replacement

on

 

 

 

4)

External power : check polarity and voltage of supply;

 

 

: check if functional on batteries.

 

 

 

 

1)

Check calibration process (as described in section 1.4).

 

2)

Check flow rate of calibrating gas (6 – 8 lpm)

 

3)

Check sensor cell connection.

 

4)

Check sensor cable connection.

Inability to achieve

5)

Check quality of calibration gas.

calibration

 

 

 

6)

Check that the calibration gas is dry (not humidified).

 

7)

Wait for 1 or 2 minutes, then attempt a re-calibration.

 

8)

Check sensor cell for damage or electrolyte leaks.

 

9)

Return monitor to authorized service agent for repair.

 

 

 

Constant non-mutable

1)

Check high and low alarm settings for possible “crossed” settings

alarm condition

 

i.e. high threshold value is set lower than low threshold value.

 

 

 

 

1)

Do not hold the sensor whilst calibrating, as the warmth from your

 

 

hand may cause the ambient temperature compensation circuit to

Drift is displayed

 

activate, resulting in unstable displayed values.

 

 

readings whilst

2)

Try calibrating a known “good” sensor.

calibrating or

 

 

immediately thereafter

3)

Repeat calibration.

 

 

4)

Replace the sensor cell.

 

 

 

There is no reaction to changes in oxygen concentration

1)Remove the sensor tip from the T-adaptor, unscrew the “flow diverter” and gently swab any excess moisture off the sensor cell membrane, then flow dry gas over the membrane.

TELEDYNE ELECTRONIC TECHNOLOGIES

Analytical Instruments

18

Image 18
Contents TED Warranty Statement All Rights ReservedTable of Contents Section Topic Number Introduction Applications Description of Functional Zones Front PanelRear Panel SET-UP Section ONE OperationsInstalling the Batteries Installing the Oxygen Sensor Calibrate the Instrument Normal Operation SET the Alarms Oxygen MonitorSection TWO Anaesthetic AgentsCleaning the TED 191 Portable Oxygen Monitor Sterilising the T-7 Oxygen Sensor and Connecting CableSection Three DO’S Section FourChanging Batteries Section FiveRoutine Visual Inspection Changing the Oxygen Sensor CellSection SIX Troubleshooting Remove the sensor tip from the T-adaptor, unscrew the flow Section Seven Technical Specifications Storage ServiceSensor Response Time TransportSection Eight Effect of PressureSection Nine HumiditySection TEN Effect Temperature Section Eleven Discrepancy Readings Section Twelve Repair Service Sensor Cell Replacement LOG Technical Service LOG General Notes Spare Parts List Optional AccessoriesSpare Parts

TED 191 specifications

The Teledyne TED 191 is an innovative device designed to enhance various technologies in the realms of data acquisition and signal processing. This cutting-edge equipment has garnered attention for its outstanding performance, robustness, and versatility, making it a popular choice among professionals in numerous industries, including aerospace, military, and environmental monitoring.

The TED 191 features a compact form factor, enabling easy integration into existing systems without requiring significant modifications or additional space. This design is especially beneficial for applications requiring portability, as the device can be seamlessly transported and deployed in the field. Moreover, its rugged construction provides durability and reliability under harsh operating conditions, ensuring consistent performance even in challenging environments.

One of the standout characteristics of the Teledyne TED 191 is its advanced signal processing capabilities. The device is equipped with high-resolution analog-to-digital converters that enable the capturing of intricate waveforms with precision and accuracy. This is crucial for applications such as seismic monitoring and geophysical surveying, where minute variations in signals can carry significant implications.

In addition to its robust data acquisition features, the TED 191 supports multiple communication protocols, allowing for seamless integration with other systems and devices. This interoperability ensures that users can easily transfer data and remotely control the device, enhancing operational efficiency. The device is compatible with various software applications, further expanding its utility for users who require specialized analytics or data visualization.

Another notable technology embedded in the TED 191 is its real-time processing capabilities. This allows users to analyze incoming data instantaneously, facilitating timely decision-making. This feature is particularly valuable in critical applications like military operations and emergency response scenarios, where seconds count, and immediate information can influence outcomes.

In summary, the Teledyne TED 191 stands out as a highly capable tool for professionals demanding high performance and reliability in data acquisition and signal processing. Its compact design, advanced signal processing technologies, interoperability, and real-time analysis capabilities make it an essential device across various sectors. As industries continue to evolve and seek out cutting-edge technologies, the TED 191 remains at the forefront, setting a standard for performance and versatility.