4.Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell, warranted for 6 months, with a life expectancy of 6 months.
5.Analog Voltage Output. The Model 319R provides customer connections to external devices, such as computers, recorders, alarms, etc., at 0–10 VDC, negative ground, plus contact closures for range identifica- tion. Custom voltage outputs, either grounded or isolated, are available as options.
6.Analog Current Output. The Model 319R has a 4–20 mA current output suitable for connection to external devices that use a 4–20 mA input.
7.Over-Range/Cell Failure Visual Indicator. The Model 319R has a LED (Light Emitting Diode) to the left of the Liquid Crystal Display. The LED will blink when the oxygen reading exceeds 105% of scale, or a cell failure alarm is activated.
Micro-Fuel Cell
The Model 319R uses a Class Z-2 Micro-Fuel Cell. The Micro-Fuel Cell measures the concentration of oxygen in a gas stream. The analysis is specific for oxygen; i.e. the measuring cell will not generate an output current unless oxygen is present in the sample gas. Therefore, the analyzer has an absolute zero, and requires no zero gas. Span calibration is per- formed with atmospheric air as the span gas.
To eliminate any inaccuracies due to the positive temperature coeffi- cient of the Micro-Fuel Cell, a thermistor circuit and network of precision resistors produce a negative coefficient of matching characteristics. The variable element (thermistor) in the compensation network is in the same assembly as the Micro-Fuel Cell, exposing both devices to the same tem- perature conditions.
Warranty Conditions
Customers having warranty claims must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty material or workmanship the cell will be replaced free of charge.
CAUTION: Evidence of tampering or abuse will render the warranty null and void.
2 | TELEDYNE BROWN ENGINEERING |