Teledyne 319R instruction manual Calibration

Page 11

319R

Installation & Operation 2.0

 

 

 

 

 

 

 

 

 

Calibration

To calibrate the 319R:

1.Place the power switch in the ON position.

2.Allow the LCD reading to stabilize.

3.Expose the sensor to ambient air (20.9%) or instrument grade air. If the analyzer goes overrange, the display will go blank and the LED to the left of the display will blink. Hold down the SPAN button until the LED stops blinking.

4.While holding down the SPAN AT 20.9% button, adjust the potentiometer to the right of the button until the display is stable and reads 20.9%.

5.Release the button. The unit is now calibrated.

6.The alarms will be disabled for about 25 seconds after the SPAN button is released. Disabling the alarms allows air to be purged from the sample system without tripping any alarm set below span, or 21%.

NOTE: Do not attempt to adjust any alarm setpoints while the alarms are disabled during the 25-second period.

TELEDYNE BROWN ENGINEERING

7

Analytical Instruments

Image 11
Contents 319R Important Notice Table of Contents Teledyne Brown Egineering Introduction Main Features319R Micro-Fuel Cell Warranty ConditionsIntroduction 319R Location & Mounting Installation & OperationPrecautions Output Signals Installation & Operation 319RAC Power International Ranges DC PowerSensor Installation or Replacement Alarm Terminal ConnectionSetting Alarms Calibration Installation & Operation 319R Replacing the Fuse 319R MaintenanceReplacing the Sensor Maintenance 319RWarranty Conditions Maintenance 319R 319R Appendix SpecificationsAppendix 319R Spare Parts ListDrawing List Section I Product Identification Section III Physical/Chemical CharacteristicsMaterial Safety Data Sheet Section VI Health Hazard Data Section IV Fire and Explosion Hazard DataSection V Reactivity Data Protective measures during cell replacement Section VII Precautions for Safe Handling and UseSection X References Section Viii Control MeasuresSection IX Disposal