necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings.
For atmospheric pressure sampling, connect a pump and flow control valve downstream from the analyzer and draw (rather than push) the sample through the instrument.
TAI supplies three male disconnect fittings with the instrument; one for installation of the customer’s sample line, one to be used to open the vent fitting of the instrument, and one (equipped with a plastic tube) for drawing air through the unit for calibration purposes.
2.2Power Service: A source of
3.OPERATION
3.1Introduction: The Model 311D is delivered completely assembled and ready for instant use. The
When the range selector is advanced from the “OFF” position, power to the instrument’s circuitry is established. The meter will instantly respond to the residual oxygen within the integral sample passages.
It is impossible to achieve a “perfect” seal of the internal sample system, and what the meter is indicating is the diffusion/consumption balance point of the internal sample system and the
This “balance” point, with a properly calibrated instrument, is always within the limits of the 100 ppm range. If the reading claims off the limits of this scale, a leak in the manifold assembly is indicated.
TO EXTEND THE CELL LIFE AND MINIMIZE THE TIME REQUIRED TO MAKE THE NEXT ANALYSIS, THE INSTRUMENT SHOULD ALWAYS BE PURGED WITH THE SAMPLE OR AN INERT GAS PRIOR TO BEING TAKEN OUT OF SERVICE FOR STANDBY OR STORAGE.
3.2Calibration: The inherently constant output of the cell during its useful life pre- cludes a definitive calibration cycle. TAI feels that the interval between calibrations should be