Teledyne 311-D instruction manual Operation

Page 6

necessary or desirable, it must have a metallic diaphragm. Regulators with organic or plastic diaphragms are permeable to oxygen, and if used in the sampling system, will lead to high oxygen readings.

For atmospheric pressure sampling, connect a pump and flow control valve downstream from the analyzer and draw (rather than push) the sample through the instrument.

TAI supplies three male disconnect fittings with the instrument; one for installation of the customer’s sample line, one to be used to open the vent fitting of the instrument, and one (equipped with a plastic tube) for drawing air through the unit for calibration purposes.

2.2Power Service: A source of single-phase, 105 to 125 volt, 50 or 60 Hertz power, capable of delivering a maximum of 1/4 ampere of current will be periodically required to recharge the instrument’s battery power supply. An eight-foot, UL approved, 3-wire, detachable power cord is provided with the instrument and should be stored in a safe place when not in use. As a no-cost option, the 311D can be furnished with 220-volt, 50 or 60 Hertz charging power.

3.OPERATION

3.1Introduction: The Model 311D is delivered completely assembled and ready for instant use. The Micro-Fuel Cell is in place within the manifold, and prior to shipment the manifold was purged with an inert gas to eliminate all but traces of oxygen from the internal sampling system. The integral shutoff valves in the quick-disconnect sample fittings, if not disturbed, will maintain this inert atmosphere within the manifold indefinitely. This can be demonstrated by advancing the range selector switch to the 0-1000 ppm position.

When the range selector is advanced from the “OFF” position, power to the instrument’s circuitry is established. The meter will instantly respond to the residual oxygen within the integral sample passages.

It is impossible to achieve a “perfect” seal of the internal sample system, and what the meter is indicating is the diffusion/consumption balance point of the internal sample system and the micro-fuel cell.

This “balance” point, with a properly calibrated instrument, is always within the limits of the 100 ppm range. If the reading claims off the limits of this scale, a leak in the manifold assembly is indicated.

-4-

TO EXTEND THE CELL LIFE AND MINIMIZE THE TIME REQUIRED TO MAKE THE NEXT ANALYSIS, THE INSTRUMENT SHOULD ALWAYS BE PURGED WITH THE SAMPLE OR AN INERT GAS PRIOR TO BEING TAKEN OUT OF SERVICE FOR STANDBY OR STORAGE.

3.2Calibration: The inherently constant output of the cell during its useful life pre- cludes a definitive calibration cycle. TAI feels that the interval between calibrations should be

Image 6
Contents For Trace Oxygen Analyzer Model 311-D Table of Contents Model 311D Introduction Page Supporting Equipment and Services Operation Be Sure to Relock the Control After the ADJUST- Ment is Made Maintenance Page Transduction and Temperature Compensation Leak Testing Specification DataOxygen Effect of CO2 on B-2C Cell Life Recommended Spare Parts List QTY DescriptionDrawing List Model 311D Page