Pioneer Q150 owner manual Troubleshooting, Humidifier, Main Power Loss, Call dealer for service

Page 16

10.4 HUMIDIFIER

NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these steps

1.Clean the humidifier every day.

2.Wash it in soapy water and rinse it with a solution of 10 parts water and 1 part vinegar.

3.Rinse again with hot water and refill with distilled water to the level shown on the humidifier.

11. TROUBLESHOOTING

 

 

 

 

 

 

 

 

 

 

 

 

SYMPTOM

 

PROBABLE CAUSE

 

SOLUTION

 

 

 

 

 

 

 

 

Alarm:

 

 

Main Power Loss:

 

 

 

 

One (1) short beep, long pause

1.

Power cord NOT plugged in.

1.

Insert plug into outlet

 

 

 

2.

No power at outlet.

2.

Inspect house circuit breakers

Concentrator NOT operating,

3.

Tripped circuit breaker.

 

or fuses. If problem recurs,

while Power switch ON

4.

Internal repairs required.

 

use a different outlet.

 

 

 

 

 

 

 

3.

Reset the circuit breaker.

 

 

 

 

 

 

 

4.

Call dealer for service.

Alarm:

 

 

System Failure:

1.

Remove and clean cabinet

Two (2) short beeps, long pause

1.Unit overheating due to blocked

 

filter.

 

 

 

air intake.

2.

Move oxygen concentrator at

 

 

 

2.

Internal repairs required.

 

least three-inches away from

Concentrator NOT operating,

 

 

 

 

 

walls, draperies or furniture.

while Power switch ON

 

 

 

 

3.

Call dealer for service.

Alarm:

 

 

System Failure:

 

 

 

 

Three (3) short beeps, long pause

1. Pneumatically the system is

1.

Shut down the system.

 

 

 

 

operating at over pressure.

2.

Call dealer for service.

 

 

 

2.

Internal repairs required.

 

 

 

 

 

 

 

 

 

 

Alarm:

 

 

System Failure:

1.

Remove and clean cabinet

Four (4) short beeps, long pause

1. Pneumatically the system is

 

filter.

 

 

 

 

operating at under pressure.

2.

Move oxygen concentrator at

 

 

 

2.

Internal repairs required.

 

least three-inches away from

 

 

 

 

 

 

 

 

walls, draperies or furniture.

 

 

 

 

 

 

 

3.

Call dealer for service.

YELLOW or RED LIGHT

1.

 

Low oxygen purity.

1.

Clean or replace filters

illuminated.

2.

 

Insufficient power at outlet.

2.

Don’t use extension cords.

 

 

 

3.

 

Internal repairs required.

 

Move to another electrical

 

 

 

 

 

 

 

 

outlet or circuit.

 

 

 

 

 

 

 

3.

Call dealer for service.

15

Image 16
Contents Pioneer Page Table of Contents IEC Symbols Page Precautions Intended USEDescription Specifications Shipping Temperature CabinetRegulatory Listing Change in Max. Flow withHandling UnpackingOperation Plug in Power CordSelect a Location Connect Oxygen Tubing Humidifier CompartmentTurn on Power Switch Adjust FlowrateOxygen Purity Indicator Start Oxygen TherapyMaintenance AlarmsGeneral Caring Cleaning the Cabinet Filters CabinetTroubleshooting Main Power LossCall dealer for service HumidifierWarranty Authorized European Representative Manufactured by