6. SPECIFICATIONS
Electrical Requirements | 115VAC±10, 60Hz / 230VAC±10, 50Hz | ||||
|
|
|
|
|
|
Rated Current Input | 3.5A (115VAC/60Hz); 2.0A (230VAC/50Hz) | ||||
|
|
|
|
|
|
Sound Level | 45dB(A) Average |
|
| ||
|
|
|
|
|
|
Operating Altitude | Do not use at altitude over 4000 meters. | ||||
|
|
|
|
|
|
Oxygen Concentration | 90at 1~5LPM |
|
| ||
| * Stated performance maximum achieved after 60 min. of run time. | ||||
Maximum Oxygen | 6±0.5psi. (40±4 kPa.) |
|
| ||
Delivery Pressure: |
|
|
|
|
|
|
|
|
|
|
|
Optional Oxygen Sensor |
|
|
|
|
|
|
| Oxygen Purity | LED Indicator |
| |
|
|
|
|
|
|
|
| Over 85(±3) | GREEN Indicator light |
| |
|
|
|
|
|
|
|
| 70(±3) to 85(±3) | YELLOW Indicator light |
| |
|
|
|
|
|
|
|
| Below 70(±3) | RED Indicator light |
| |
|
|
|
|
|
|
|
|
|
|
| |
Flow Range / Maximum |
|
| |||
Recommended Flow |
|
|
|
|
|
|
|
|
|
| |
Average Power | 320W |
|
| ||
Consumption |
|
|
|
|
|
|
|
|
|
| |
Pressure Relief Activation | 250kPa±10 |
|
| ||
|
|
| |||
Filters: | Cabinet, Bacteria and Compressor Inlet | ||||
|
|
| |||
Alarms – (Visual/Audible) | High temperature w/ compressor – power shutdown | ||||
| High / Low pressure alarm |
|
| ||
| Power loss alarm |
|
| ||
|
|
| |||
Dimensions | 68cm(26.8”) High × 45cm(17.7”) Wide × 33cm(13”) Deep | ||||
|
|
|
| ||
Circuit Breaker | (115VAC/60Hz) |
| (230VAC/50Hz) | ||
& Fuse | Circuit Breaker: 8A |
| Circuit Breaker: 4A | ||
Weight | 28.2Kg (62lb) |
|
| ||
|
|
|
| ||
Shipping Weight | 31.8Kg (70lb) |
|
| ||
|
|
|
| ||
Operating Ambient | 10 ~ 39 |
|
| ||
Temperature |
|
|
|
|
|
|
|
|
|
|
|
7