Omron CX Pro instruction manual Cleaning and disinfection, Storage and maintenance

Page 8

English

Cleaning and disinfection

Disinfect the nebuliser before first use and at least once per day when in use. Procedure:

Disassemble the nebuliser kit by disconnecting the air tube and mouthpiece or mask. Remove the nebuliser cap from the medication tank and remove the vaporizer head. Also refer to the illustrations as shown under step 3 and 4 (page 5).

Disinfect all disassembled nebuliser parts –NOT the air tube and masks- by placing them in boiling water (max.10 minutes) or in an autoclave (max. 121ºC, refer to the autoclave instructions for the duration).

To clean the nebuliser and mask immediately after each treatment, it is sufficient to rinse the disassembled parts under running warm tap water.

The masks can be disinfected by any commercially available disinfectant, containing hypochloride. After cleaning/disinfecting of the parts, place them on a clean (paper) towel and let them air dry.

For hygienic reasons, each person should use their own nebuliser kit. It is recommended that the nebuliser kit is replaced every 6 months (or earlier in case the nebuliser is obstructed) in order to maintain efficient inhalation. Contact your OMRON dealer for replacements.

The exterior of both the air tube and compressor unit should be cleaned once daily with a clean, damp piece of cloth. Do not use any volatile fluids. Never submerge the compressor unit in water!

If there is some moisture in the air tube, connect the air tube to the compressor unit and leave the other end open. Then switch on the compressor unit and blow the tube dry.

The exterior of both the air tube and compressor unit should be cleaned once daily with a clean, damp piece of cloth. Do not use any volatile fluids. Never submerge the compressor unit in water!

If there is some moisture in the air tube, connect the air tube to the compressor unit and leave the other end open. Then switch on the compressor unit and blow the tube dry.

Storage and maintenance

Put the nebuliser kit, masks and airtube in a clean bag and place them in the storage pocket of the car- rying case.

Wrap the mains cable and place it in its storage compartment at the rear of the compressor unit.

Do not store the compressor on the floor.

The unit does not contain any user serviceable parts and does not require oil. Do not attempt to lubricate any internal parts. Do not open the compressor unit.

In case you encounter difficulties with your nebuliser, please contact your local OMRON dealer. (address on/inside package).

Disposal of this product should be carried out in accordance with the national regulations for the disposal of electronic products.

7

Image 8
Contents CX PRO Table of Contents Dear CustomerSafety precautions General InformationDescription of the Omron CompAIR CXpro Nebuliser Description of the nebuliser kit JetAIR plus Inhaling with the Omron CompAIR CX pro Nebuliser Breathe out Cleaning and disinfection Storage and maintenanceFilter Possible problems and how to solve themOptional accessories/ parts Technical data Output Rate0197 Omron Healthcare Europe B.V. IM-NE-C18-E-02-07/07