ResMed S8 Elite II manual Troubleshooting, No display

Page 14

Please consult the medical services department of your carrier if you intend to use your S8 Series II on an aircraft. The S8 Series II complies with US FCC Part 15, Class B requirements.

Note: Do not use your S8 Series II while the aircraft is taking off or landing.

Can I use my S8 Series II if mains (AC) power is not available?

A DC-12 converter is available to connect the S8 Series II to a 12V or 24V DC power source. Contact your clinician for details.

Troubleshooting

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the device.

Problem/Possible Cause

Solution

 

 

No display

 

Power is not connected.

Ensure the power cable is connected and the power

 

switch (if available) is on.

Insufficient air delivered from S8 Series II device

Ramp or settling time is in use.

Wait for air pressure to build up or change ramp or settling time.

Air filter is dirty.

Air tubing is kinked or punctured.

Air tubing not connected properly.

Mask and headgear not positioned correctly.

Replace air filter.

Straighten or replace tubing.

Check air tubing.

Adjust position of mask and headgear.

Cap(s) missing from access port on

Replace cap(s).

mask.

 

Pressure required for treatment may

See your clinician to adjust the pressure.

have changed.

 

Device does not start when you breathe into the mask although your clinician has enabled SmartStart

Power cord not connected properly.

Power outlet may be faulty.

Leak Alert has been enabled by your clinician; SmartStart automatically disabled.

Use of a ResMed Full Face mask.

Breath not deep enough to trigger SmartStart.

Connect power cord firmly at both ends.

Try another power outlet.

Contact your clinician.

SmartStart does not work with a full-face mask as the anti-asphyxia valve will not allow sufficiently high pressure on exhalation.

Take a deep breath in and out through the mask.

12

Image 14
Contents S8 Series Contents Adverse Effects IntroductionContraindications Parts of the S8 Series II system Masks Humidifiers ResScan Data CardUsing the S8 Series Setting UpHow to Use the Control Panel Selecting the Mask TypeUsing the Mask-fitting Feature Starting TreatmentStandard Menu Changing the Settings on your S8 II DeviceStopping Treatment Detailed Menu Cpap AutosetCopying Data onto the Data Card Using the Data Card to Collect DataMessages on the S8 Series II LCD Weekly Cleaning and MaintenanceDaily MonthlyReplacing the Air Filter ServicingFrequently Asked Questions How will it feel when I am breathing with the S8 Series II?Troubleshooting No displayDisplays error message Check tube!! Key if done Technical Specifications Power SupplySymbols which appear on the device General Warnings and CautionsPage Limited Warranty

S8 Elite II specifications

The ResMed S8 Elite II is a continuous positive airway pressure (CPAP) device designed to provide effective treatment for obstructive sleep apnea (OSA). Renowned for its reliability and efficiency, this model brings together advanced features and user-friendly technology aimed at enhancing the quality of sleep for individuals suffering from this condition.

One of the most notable features of the S8 Elite II is its AutoSet technology, which ensures that the device adjusts the airflow pressure automatically. This allows for optimal treatment pressure during sleep, responding to the varying needs of the patient throughout the night. Consequently, patients experience a more comfortable and less intrusive therapy session as the device adapts to changes in their breathing patterns.

The S8 Elite II's Easy-Breathe technology is another standout characteristic. This innovative feature mimics the natural breathing pattern, providing a more comfortable experience during inhalation and exhalation. As a result, users report decreased air pressure discomfort, making it easier to acclimate to their CPAP therapy.

In terms of usability, the S8 Elite II is designed with simplicity in mind. The intuitive menu and adjustable settings allow users to personalize their therapy easily. The large LCD screen provides clear visual feedback and guidance, helping patients to make quick adjustments as needed. Additionally, it comes with an integrated SmartStart function, which begins therapy automatically when the mask is placed on, and pauses when it is removed. This seamless operation enhances the user experience and encourages consistent usage.

Portability is another significant advantage of the ResMed S8 Elite II. Its compact design allows for easy transportation, making it ideal for travel. Users appreciate its lightweight frame and sturdy carry bag, ensuring that essential sleep therapy is never compromised, even on the go.

Moreover, the device's data tracking capabilities enable patients and healthcare providers to monitor therapy effectiveness. The built-in data card stores nightly usage information, which can be accessed and analyzed during follow-up appointments to optimize treatment.

In summary, the ResMed S8 Elite II stands out in the field of sleep apnea therapy devices. Its advanced technologies, such as AutoSet and Easy-Breathe, make it a top choice for users seeking effective and comfortable treatment. With its user-friendly design, portability, and data tracking features, the S8 Elite II is equipped to enhance sleep quality and overall well-being for those living with obstructive sleep apnea.